Why it was recalled

GE Healthcare has recently become aware of a potential safety issue due to the interlock slide mechanism missing from the 3-vaporizer manifold of your Aestiva 7900 Anesthesia Device. This Potential for two (2) vaporizers delivering agent at the same time. This could manifest as over delivery of a single agent if both vaporizers contain the same agent or as delivery of more than one agent. This m

FDA root cause category: Process control

What the company is doing

GE Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated April 10, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Contact GE Healthcare Customer or Technical Support at 800-345-2700 for questions regarding this notification.

Recall details

Recall number
Z-1787-2012
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Apr 10, 2012
Posted by the FDA
Jun 17, 2012
Recalling firm
GE Healthcare, LLC
Quantity
8
Product code
CBK
FDA submissions
K023366
Distribution
Nationwide Distribution - including the states of AL, FL, GA, MO, and VA.

Product description

GE Healthcare, Aestiva/5 7900 SmartVent, anesthesia machine. Used in Datex-Ohmeda Aestiva/5 Anesthesia Systems.

Code information (models, lots, serial numbers)

Serial Number AMRP01031 AMRP00966 AMRP01030 AMRP00968 AMRP00967 AMRP01033 AMRP00970 AMRP00969

Other devices under the same FDA submission

1 device cleared under K023366 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates