CIVCO General Purpose Sensor, Non-sterile reusable general… recallZ-1954-2012 · Civco Medical Instruments Inc · initiated May 7, 2012
Class II Terminated All recalls posted in July 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Civco Medical Instruments Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device.
FDA root cause category: Component design/selection
What the company is doing
CIVCO sent an Urgent Medical Device Recall letter dated May 3, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to return all general purpose sensors as the product may not function as intended. To return the affected product customers should contact the OEM Sales Coordinator at 319-248-6521 to schedule the return of their device. Upon receipt of the snesor, CIVCO will arrange for recalibration. Once recalibration is completed, the sensor will be returned to the customer at no charge. Customers were asked to forward this communication to all their customers who purchased the affected product. Once CIVCO has received the customer list, they will contact the end-user to request they return their general purpose sensor directly to CIVCO for recalibration. For questions customers should call 800-441-6741. For questions regarding this recall call 319-656-4447.
Recall details
- Recall number
- Z-1954-2012
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 7, 2012
- Posted by the FDA
- Jul 6, 2012
- Recalling firm
- Civco Medical Instruments Inc
- Quantity
- 100 kits
- Product code
- IYO
- FDA submissions
- K092619
- Distribution
- Worldwide Distribution - USA including WI, CA, MA, IN, WA, and NJ. Foreign distribution was made to Germany, Canada, Denmark, Hong Kong, Ukraine, Japan, Italy, Korea, and France.
Product description
CIVCO General Purpose Sensor, Non-sterile reusable general purpose electromagnetic sensor for use with 3D Guidance driveBAY, driveBAY 2, trakSTAR, or trakSTAR2 trackers, Ref. 610-1066. The firm name on the label is CIVCO, Kalona, IA. The kit contains sensor component part number 410-240. The general purpose sensor within the kit is intended to provide physicians with tools for electromagnetic tracking of instruments with respect to image data. The sensors are designed to attach to custom reusable tracking bracket for leading transducers and other medical devices
Code information (models, lots, serial numbers)
Lot numbers M436540, M468460, M469640, M495160, M495190, M523420, M524580, M555880, M563330, M577940, M590400, M597880, M643870, M673080, M685650, M708280, M717280, M726110, M726130, M736410, M782740, M799860, M817860, and M840610.
Other devices under the same FDA submission
12 devices cleared under K092619 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VirtuTRAX Instrument Navigator 5350925Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- VirtuTRAX Instrument Navigator 610-1080Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- eTRAX™ Needle System Starter Kit 667-149Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- eTRAX™ Needle System Starter Kit 667-150Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- eTRAX™ Needle System Starter Kit 667-151Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- eTRAX™ Needle System Starter Kit 667-152Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- eTRAX™ Needle Sensor 667-156Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- eTRAX™ Needle Sensor 667-157Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- eTRAX™ Needle Sensor 667-158Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- eTRAX™ Needle Sensor 667-159Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- VirtuTRAX Instrument Navigator 610-1059Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- VirtuTRAX Instrument Navigator 610-1153Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 61793).
- Z-1952-2012 · TerminatedCIVCO VirtuTRAX Instrument Navigator, Reusable non-sterile general purpose driveBAY, trakSTAR, electromagnetic sensor (8mm) with VirtuTRAX instrument navigator and (8.9 tapered to …
- Z-1953-2012 · TerminatedCIVCO General Purpose Sensor, Non-sterile reusable general purpose electromagnetic sensor for use with 3D Guidance driveBAY, driveBAY 2, trakSTAR, or trakSTAR2 trackers, Ref. 10441…
- Z-1955-2012 · TerminatedCIVCO General Purpose Sensor, Non-sterile reusable general purpose electromagnetic sensor, for use with 3D Guidance medSAFE trackers, Ref. 610-1067. The firm name on the label is…
- Z-1956-2012 · TerminatedCIVCO VirtuTRAX Instrument Navigator, Reusable non-sterile general purpose driveBAY, trakSTAR, electromagnetic sensor (8mm) with VirtuTRAX Instrument navigator and (8.9 tapered to …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page