Abbott Molecular Bar Code Scanner User's Guide recallZ-2013-2012 · Abbott Molecular · initiated Jan 31, 2012
Class II Terminated All recalls posted in July 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Molecular or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
In the Abbott Molecular Bar Code Scanner User's Guide, List Number 6L88-02/09, the section Modifying the Keyboard Format for Bar Code Scanner LN 6L89-01 contains incorrect bar codes that can prevent the correct configuration of the bar code scanner language format. The Abbott m2000 System is intended for use in performing nucleic acid testing in clinical laboratories. It is comprised of the Ab
FDA root cause category: Error in labeling
What the company is doing
Abbott Molecular Inc. sent an Urgent Field Safety Notice letter dated January 31, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customer was instructed to replace the affected pages of the Bar Code Scanner User's Guide with the replacement pages provided with the letter, and immediately perform the Modifying the Keyboard Format section with these replacement pages. The accounts were requested to complete the enclosed reply form and fax it to (847) 775-6728 to acknowledge receipt, understanding and implementation of the information provided in the notice.
Recall details
- Recall number
- Z-2013-2012
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jan 31, 2012
- Posted by the FDA
- Jul 14, 2012
- Recalling firm
- Abbott Molecular
- Quantity
- 100 units
- Product code
- JQW
- FDA submissions
- K092705
- Distribution
- California, Florida, Maryland, Oregon and Virginia, and internationally to Belgium, Brazil, China, France, Germany, Hong Kong, Japan, Malaysia, Mexico, Qatar, Switzerland and United Kingdom
Product description
Abbott Molecular Bar Code Scanner User's Guide; instructions for setting up the Bar Code Scanner LN 6L89-01, a component of the Abbott m2000 Instrument System; Abbott Molecular Inc., Des Plaines, IL 60018 USA; list number 06L88-02. Language configuration bar codes - Intended Use: The bar code scanner is a hand-held device that provides a convenient means of scanning sample bar codes to allow positive sample identification and is supplementary to the computer keyboard. Both the scanner and the keyboard transmit data to a System Control Center (SCC) through a single wedge (cable with branched connectors). The bar code scanner must be configured to match the keyboard language being used. In its default configuration set by the manufacturer (Jadak), the scanner is in English format. If a keyboard other than the English keyboard is being used, the user is required to use the scanner to scan an appropriate language configuration bar code provided in the User's Guide to match the language of the keyboard. Users in the U.S have the ability to convert to another language.
Code information (models, lots, serial numbers)
list numbers 06L88-02, 06L88-03, 06L88-04, 06L88-05, 06L88-06, 06L88-07, 06L88-08 and 06L88-09, all lots
Other devices under the same FDA submission
2 devices cleared under K092705 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- m2000 09K14-02Abbott Molecular Inc. · CLINICAL SAMPLE CONCENTRATOR23 reports
- m2000 9K15-01Abbott Molecular Inc. · real time Nucleic acid amplification system6 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.