I-STAT PT/INR cartridge Abbott Point of Care Inc. Abbott… recallZ-2157-2012 · Abbott Point Of Care Inc. · initiated Mar 1, 2012

Class II Terminated All recalls posted in August 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Point Of Care Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The i-STAT PT/INR cartridges have the potential to exhibit incorrectly elevated results.

FDA root cause category: Process control

What the company is doing

Abbott Point of Care Inc, sent a URGENT RECALL NOTICE dated March 2012 , to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were provided in the letter with instructions in the letter on how to handle the affected product. Please complete and return the enclosed Business Reply card to acknowledge receipt of this communication and installation of the interium software CLEW and also 1) Update all handhelds 2) download appropriate assignment sheets. For further questions please call (609) 454-9000.

Recall details

Recall number
Z-2157-2012
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 1, 2012
Posted by the FDA
Aug 8, 2012
Recalling firm
Abbott Point Of Care Inc.
Quantity
2,950,296 cartridges US; 156,408 cartridges OUS
Product code
GJS
FDA submissions
K020355
Distribution
Worldwide Distribution -- (USA) nationwide

Product description

I-STAT PT/INR cartridge Abbott Point of Care Inc. Abbott Park, IL 60064 USA This cartridge is useful for monitoring patients receiving oral anticoagulation therapy such as Coumadin or warfarin.

Code information (models, lots, serial numbers)

List number 04J50-01 04J5O-02 03P89-24  Lots begininng with R11 or R12

Other devices under the same FDA submission

1 device cleared under K020355 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.