HALO Flex Energy Generators recallZ-2379-2012 · Covidien LLC · initiated Aug 3, 2012
Class II Terminated All recalls posted in September 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential failure of HALO Energy Generator to enter "stand by" mode when initially powered on; and therefore can not be used for patient care.
FDA root cause category: Software design
What the company is doing
COVIDIEN sent an Urgent Medical Device Correction letter dated August 3, 2012, to all affected customers via Federal Express. Additionally, Field Service Representatives will be contacting their accounts to notify them of the correction.. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to take the following steps prior to initiating patient treatment and specifically, prior to administering conscious sedation and performing upper endoscopy: 1. Power-on the HALO Flex Energy Generator and confirm that the generator passes self-test, enters standby mode and displays "Ready, Connect Catheter." 2. Confirm that the HALO Flex Energy Generator is functioning properly at power-up and enters the standby mode then the procedure can begin. 3. Do not use the HALO Flex Energy Generator if the generator does not pass the power on self-test. In this situation, please call Covidien, GI Solutions customer service at (888) 662-2779 or your local territory manager (sales representative) so that the firm can provide them with a replacemenent generator. 4. Customers were asked to reply to Covidien, GI Solutions (formerly BARRX Medical) by signing and returning the attached Product Correction Form via fax to the number provided on the form. Customers with questions were instructed to call 888-662-2779. For questions regarding this recall call 408-328-7300.
Recall details
- Recall number
- Z-2379-2012
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Aug 3, 2012
- Posted by the FDA
- Sep 14, 2012
- Recalling firm
- Covidien LLC
- Quantity
- 617 generators
- Product code
- GEI
- FDA submissions
- K092487
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia, Austria, Belgium, Canada, Chile, Colombia, Czech Republic, Denmark, France, Germany, Greece, Hungary, India, Israel, Jordan, Netherlands, Norway, Poland, Portugal, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, Venezuela and UK.
Product description
HALO Flex Energy Generators Manufactured by BRRX Medical (now Covidien) 540 Oakmead Parkway Sunnyvale, CA 95085. The HALO Flex Energy Generator is indicated for use for the coagulation of soft tissue. The HALO Flex System is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to, the esophagus, including Barrett's Esophagus.
Code information (models, lots, serial numbers)
Model numbers 1190A-115A and 1190A-230A; all serial numbers.
Other devices under the same FDA submission
1 device cleared under K092487 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Barrx 1190A-115ACovidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES11 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.