STERRAD NX Cassette, Batch # 12A078 recallZ-2338-2012 · Advanced Sterilization Products · initiated Jun 19, 2012
Class III Terminated All recalls posted in September 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Advanced Sterilization Products or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Advanced Sterilization Products (ASP) is recalling (removing) the Sterrad 100S and Sterrad NX System cassettes because the camera used to verify printing on the Sterrad System cassettes was inadvertently disabled. The STERRAD 100S and STERRAD NX System cases were shipped without verification of proper printing.
FDA root cause category: Error in labeling
What the company is doing
Advanced Sterilization Products sent an "URGENT MEDICAL DEVICE RECALL (REMOVAL) letter dated July 16, 2012 to all their customers who purchased the STERRAD 100S and STERRAD NX System cassettes. The letter informed the customers of the problem identified and the actions to be taken. For the STERRAD NX System cassettes, there should be no impact to the functionality of these affected cassettes. Although printing may be smeared, illegible, or missing, the packaging has multiple references to the information, and the cassettes should be usable despite any printing error. For the STERRAD 100S System cassettes, there is a chance that the barcode will be unreadable and the STERRAD 100S Sterilizer will not accept the cassette. Customers were instructed to complete and return the business reply card and packing slip that are included in the letter. Customers were instructed to contact Stericycle at (888) 406-9303 for additional labels. Customers with questions about the recall letter were instructed to contact Stericycle at (888) 406-9303.
Recall details
- Recall number
- Z-2338-2012
- Recall class
- Class III
- Status
- Terminated
- Initiated by the company
- Jun 19, 2012
- Posted by the FDA
- Sep 7, 2012
- Recalling firm
- Advanced Sterilization Products
- Quantity
- 7,550 units
- Product code
- MLR
- FDA submissions
- K042116
- Distribution
- Nationwide Distribution.
Product description
STERRAD NX Cassette, Batch # 12A078. The STERRAD 100S and STERRAD NX sterilizations systems are low-temperature general purpose sterilizers used to sterilize heat and moisture sensitive reusable medical devices. The sterilant used is hydrogen peroxide supplied in cassette form as a separate accessory. The STERRAD Cassettes are used in conjunction with the STERRAD Sterilizers, and each cassette model is for use only in the matching sterilizer model.
Code information (models, lots, serial numbers)
Product Code: 10133
Other devices under the same FDA submission
1 device cleared under K042116 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- STERRAD NX™ 10033Advanced Sterilization Products Services Inc. · Sterilizer, chemical132 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 62925).
- Z-2337-2012 · TerminatedSTERRAD 100S Cassette, Batch # 12A077. The STERRAD 100S and STERRAD NX sterilizations systems are low-temperature general purpose sterilizers used to sterilize heat and moisture s…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page