ADVIA Centaur FSH Assay: ADVIA Centaur Calibrator B recallZ-2410-2012 · Siemens Healthcare Diagnostics, Inc · initiated Aug 7, 2012

Class III Terminated All recalls posted in September 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

ADVIA Centaur Prolactin Assay and FSH, Calibrator B fails the analytical sensitivity stated in the IFU for lots below 40

FDA root cause category: Nonconforming Material/Component

What the company is doing

Siemens Healthcare issued an Urgent Field Safety Notice letter dated August 2012 by e-mail to all affected Siemens Healthcare Diagnostics Branch offices on August 7, 2012. Customers were instructed to discontinue using the ADVIA Centaur Calibrator B lots 40 and below following their obtaining Calibrator 8 lot 41 or higher. Customers were asked to complete and fax-back the response form attached.

Recall details

Recall number
Z-2410-2012
Recall class
Class III
Status
Terminated
Initiated by the company
Aug 7, 2012
Posted by the FDA
Sep 19, 2012
Recalling firm
Siemens Healthcare Diagnostics, Inc
Quantity
15,445 kits
Product code
CGJ
FDA submissions
K904714
Distribution
Worldwide Distribution - US Nationwide and the countries of: Canada, Argentina, Australia, Bangaldesh,Brazil, Chile, China, Columbia, Croatia, Ecuduar, Egypt, Hong Kong, India, Israel,Japan, Korea, Maylasia, New Zealand, Mexico, Peru, Philipines, Pakistan, Saudia Arabia, Singapore, Thailand, Taiwan, Viet Nam, South Africa and UK.

Product description

ADVIA Centaur FSH Assay: ADVIA Centaur Calibrator B US: Calibrator B 00652707 (6 pack) or 00649625 (2 pack ) OUS:OUS: 2 pack REF: 00650577 and Ref: 00654130 (6 pack) Product Usage: ADVIA Centaur CP FSH assay: For in vitro diagnostic use in the quantitative determination of follicle-stimulating hormone (FSH) in serum using the ADVIA Centaur and ADVIA Centaur XP systems

Code information (models, lots, serial numbers)

US: 2 Pack Ref: 00649625: Lot Numbers/Exp Date: Dec 2, 2012: 78813A30 79982A30 81013A30 82021B30 82767B30 83453C30 85447D30 87319A40 Exp Date:April 9, 2013 85598A40 85708A40 86098A40 87319A40 87979B40 88809B40 89351C40 91750C40 (6 pack) Ref. 00652707: Lot Numbers/Exp Date: Dec 2 2012 79354A30 80802A30 81892B30 82815B30 83741D30 84186A40 Exp. Date April, 8, 2013 85599A40 86191A40 88102B40 89717C40 91383C40 (6 Pack) Ref: 00654254  79714A30 Exp Date Dec 2, 2012 82370A30 84274A30 85512E30 OUS: 2 pack REF: 00650577  78319A29 Exp Dec 2, 2102 78487A29 79024A29 79403A29 79770B29 79981B29 82935B29 85735C29 00654130  Ref: 00654130 (6 pack) 78321A29 Exp Dec 2,2102 79197A29 83314C29

Other devices under the same FDA submission

2 devices cleared under K904714 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 62897).