Skip to content

Multipolar Bipolar Cup Liner 22 mm LD recallZ-0410-2013 · Zimmer, Inc. · initiated Aug 31, 2012

Class II Terminated All recalls posted in November 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Zimmer Inc. is initiating a removal of the Multipolar Bipolar Cup Liner 22 mm J.D. (item 00-5001-040-22) due to reports that the split poly ring was incorrectly assembled.

FDA root cause category: Process design

What the company is doing

Zimmer sent an Urgent Medical Device Recall letter to all affected consignees during the week of September 3, 2012. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory immediately, quarantine the recalled product and return remaining inventory. Customers were advised to complete Inventory Return Certification form attached.

Recall details

Recall number
Z-0410-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Aug 31, 2012
Posted by the FDA
Nov 20, 2012
Recalling firm
Zimmer, Inc.
Quantity
195
Product code
KWY
FDA submissions
K833991
Distribution
Worldwide Distribution - USA nationwide including the states of: AK, CA, IL, FL, HI, MI, MO, NC, NJ, NY, OH, PA, TX and the countries of Germany, Brazil, Canada, Japan, Taiwan, Thailand, India and Korea.

Product description

Multipolar Bipolar Cup Liner 22 mm LD.; Item 00-5001-040-22 Product Usage: The liner snaps into the Multipolar Bipolar Cup and serves as the articulation surface for the femoral head. The split poly locking ring helps contain the head within the liner and also provides a popping sound as conformation that the femoral head has completed seated, as noted in surgical technique 97-5003-003-00

Code information (models, lots, serial numbers)

Item 00-5001-040- 22; Lot 61261192, 61261194, 61296091.

Other devices under the same FDA submission

23 devices cleared under K833991 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates