Skip to content

Spine Frontier Inspan compressors recallZ-0412-2013 · SpineFrontier, Inc. · initiated Aug 28, 2012

Class II Terminated All recalls posted in November 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from SpineFrontier, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Inspan Compressors may break at weld causing a loss of compression

FDA root cause category: Device Design

What the company is doing

Spine Frontier issued e-mail notices to customers on May 14, 2012 advising of the potential for breakage. Units were replaced with weld free design.

Recall details

Recall number
Z-0412-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Aug 28, 2012
Posted by the FDA
Nov 21, 2012
Recalling firm
SpineFrontier, Inc.
Quantity
17 units
Product code
KWP
FDA submissions
K093438
Distribution
Worldwide Distribution-USA (nationwide) including PR and the states of FL, MA, MO, OK, and TX, and the country of Mexico.

Product description

Spine Frontier Inspan compressors Part Number: 11-60004. Inspan compressors provide pressure to drive spikes of Inspan implants into the spinous process bone, resulting in fixation of the implant.

Code information (models, lots, serial numbers)

Lot Numbers: SFP02323, SFP02323R, and SFP02323RR

Other devices under the same FDA submission

1 device cleared under K093438 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page