CM Flowmeter Pkg H, Model #36700, Serial # recallZ-0019-2013 · Accutron Inc · initiated Apr 29, 2012
Class I Terminated All recalls posted in October 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Accutron Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Accutron is recalling the Flowmeter (Analgesia Gas Machine) because when it flows Nitrous Gas without Oxygen gas being turned on, it functions in an incorrect manner.
FDA root cause category: Device Design
What the company is doing
Accutron sent an Urgent Advisory Notice letter dated April 6, 2012 to all their customers who purchased the Ultra PC% Cabinet Mount Flowmeter. The letter informed the customers of the problem identified and actions to be taken. Customers were instructed to discontinue use or to continue use of the device only with extreme care until a replacement can be installed. Extreme care includes the use of the titration method for nitrous oxide administration coupled with close monitoring of both the patient and the actual gas flow through the flowtubes. For question contact Accutron at (800) 531-2221 or by email at [email protected].
Recall details
- Recall number
- Z-0019-2013
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Apr 29, 2012
- Posted by the FDA
- Oct 9, 2012
- Recalling firm
- Accutron Inc
- Quantity
- 220 units total
- Product code
- BSZ
- FDA submissions
- K052335
- Distribution
- US Nationwide Distribution
Product description
CM Flowmeter Pkg H, Model #36700, Serial # 13749 Product Usage: Conscious Sedation Unit Analgesia Gas Machine.
Code information (models, lots, serial numbers)
Lot # 75285
Other devices under the same FDA submission
8 devices cleared under K052335 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Flowmeter, Digital Flushmount 27076Accutron, Inc. · Gas-Machine, Anesthesia7 reports
- Digital Ultra Flushmount Package H 56800Accutron, Inc. · Gas-Machine, Anesthesia2 reports
- Digital Ultra Flushmount Package H CHAIR 56800-CHAIRAccutron, Inc. · Gas-Machine, Anesthesia2 reports
- Newport Flowmeter System 50000Accutron, Inc. · Gas-Machine, Anesthesia1 report
- Digital Ultra Package B 56100Accutron, Inc. · Gas-Machine, Anesthesia1 report
- Flowmeter, Digital Flexmount, Australian 51975-AUSAccutron, Inc. · Gas-Machine, Anesthesia1 report
- Flowmeter, Digital FlexMount, Wall Arm 26400-WAAccutron, Inc. · Gas-Machine, Anesthesia1 report
- Automatic Manifold, Digi-Flo 27497-FRUAccutron, Inc. · Gas-Machine, Anesthesia1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 63174).
- Z-0017-2013 · TerminatedFlow Meter, CM Assembly, Model #27025, Serial # 13686, 13690, 13712, 13793. Flowmeter, CM Assembly, Model#31970, Serial#13692. Product Usage: Conscious Sedation Unit Analgesia …
- Z-0018-2013 · TerminatedUltra PC% Cabinet Mount Package G, Model #36600, Serial # 13593, 13594, 13595, 13596, 13597, 13598, 13599, 13600, 13602, 13605, 13610, 13611, 13612, 13615, 13616, 13617, 13621, 13…
- Z-0020-2013 · TerminatedNewport Portable Package, Model #50000, Serial # 13591, 13592, 13604, 13606, 13614, 13662, 13663, 13664, 13726, 13731, 13733, 13736, 13737, 13738, 13740, 13768, 13770, 13771, 1377…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page