Skip to content

Siemens Magnetom Skyra, Verio, Spectra and Biograph MMR recallZ-2464-2012 · Siemens Medical Solutions USA, Inc · initiated Sep 12, 2012

Class II Terminated All recalls posted in September 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When positioning the Flex Large 4 Coil (part No. 8625761) off-center in left-right direction as it is used for example for hip and shoulder imaging and, in addition at an off-center position with respect to the head-foot direction, the coil can heat up in the area of the electronic housing (white plastic lids).

FDA root cause category: Device Design

What the company is doing

Siemens Healthcare sent a Customer Safety Advisory Notice dated September 5, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The letter included updated instruction to all customers. For questions regarding this recall call 610-219-6300.

Recall details

Recall number
Z-2464-2012
Recall class
Class II
Status
Terminated
Initiated by the company
Sep 12, 2012
Posted by the FDA
Sep 26, 2012
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
370
Product code
LNH
FDA submissions
K072237, K101347, K103429, K121160
Distribution
Nationwide Distribution - USA including Washington, DC and AL, AK, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI and Puerto Rico.

Product description

Siemens Magnetom Skyra, Verio, Spectra and Biograph MMR Nuclear Magnetic Resonance Imaging System and Tomographic Imager

Code information (models, lots, serial numbers)

Model numbers: 10276755 (Verio), 10684333 (Verio Dot), 10684334 (Verio Dot upgrade), 10432915 (Skyra), 10655588 (Spectra), and 10433372 (Biograph mMR)

Other devices under the same FDA submission

1 device cleared under K072237, K101347, K103429, K121160 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page