GE Discovery NM/CT670 recallZ-0121-2013 · GE Healthcare, LLC · initiated Oct 12, 2012
Class II Terminated All recalls posted in October 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE Healthcare has recently become aware of a potential safety issue associated with the table pallet of your Discovery NM/CT 670 and/or Discovery NM630 system which may impact operation safety. At scan setup, during gantry rotation towards scanning position, a potential contact between the table-pallet-edge and the detector was identified. Such occurrence may damage the pallet front-edge and
FDA root cause category: Software design
What the company is doing
GE Healthcare, LLC sent an " Urgent Medical Device Correction" letter dated October 12, 2012 to all affected customers. The letter was addressed to Managers of Nuclear Medicine, Hospital Administrators / Risk Mangers, and Mangers of Radiology / Cardiology. The letter described the Safety Issue, Safety Instructions, Affected Product Details and Product Correction. For questions or concerns contact your local GE Healthcare Service Representative. You may also contact GE Healthcare Call Center: United States or Canada: 800-437-1171 and Other countries: 262-896-2890.
Recall details
- Recall number
- Z-0121-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 12, 2012
- Posted by the FDA
- Oct 23, 2012
- Recalling firm
- GE Healthcare, LLC
- Quantity
- 210
- Product code
- KPS
- FDA submissions
- K093514, K111445
- Distribution
- Worldwide distribution: USA (Nationwide) in the states of: AL, CA, FL, GA, IL, IN, KS, LA, MD, MA, MN, MS, MO, NJ, NY, NC, OH, OK, PA, SC, TN, TX, VA, WI and WV. AUSTRALIA, BELGIUM, CANADA, CHINA, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GUADELOUPE, HONG KONG, INDIA, IRELAND, ISRAEL,ITALY, JAPAN, REPUBLIC OF KOREA, NETHERLANDS,NORWAY, POLAND, RUSSIAN FEDERATION, SPAIN, SWITZERLAND, TAIWAN,THAILAND, TURKEY, UNITED KINGDOM.
Product description
GE Discovery NM/CT670; GE Discovery NM 630 Product Usage: a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce: NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Mfg Lot or Serial # 21058, 21002, 21013, 21017, 21020, 21029, 21039, 21041, 21049, 21066, 21075, 21079, 21080, 21081, 21082, 21119, 21140, 21144, 21145, 21147, 21150, 96BH8, 21001, 21008, 21034, 21050, 21056, 21104, 21132, 22003, 22020, 22038, 210088, 21042, 21048, 21061, 21077, 21085, 21094, 21115, 21122, 21123, 21137, 21143, 22051 21005, 21072, 21028, 21052, 21076, 21083, 21092, 21101, 21113, 21118, 21131, 21138, 21024, 22001, 22011, 22023, 21019, 21055, 21154, 83BH6 /000, 21007, 21047, 21009, 21021, 21022, 21033, 21035, 21037, 21063, 21064, 21067, 21070, 21073, 21086, 21095, 21121, 21128, 21134, 21141, 21146, 21152, 21153, 22014, 22016, 21003, 21006, 21010, 21027, 21036, 21043, 21046, 21051, 21084, 21108, 21116, 21120, 21124, 21133, 22015, 22028, 21093, 21158, 22021, 21031, 21129, 00ME1, 21004, 21014, 21053, 22005, 21012, 21059, 21151, 22022, 22031, 22035, 21089, 21100, 21136, 22044, 22071, 81BH0 /000, 82BH8 /000, 92BH7 /000, 93BH5 /000, 94BH3, 21016, 21040, 21060, 21103, 21125, 22037, 21038, 22008, 21068, 21126, 22009, 21026, 21011, 21018, 21025, 21071, 21090, 21109, 21110, 21130, 21142, 21160, 21096, 21107, 21091, 21099, 21000, 21023, 21030, 21032, 21044, 21045, 21054, 21062, 21065, 21069, 21074, 21105, 21111, 21114, 21127, 21149, 22002, 22004, 22007, 22010, 22012, 22017, 22018, 22024, 22025, 22026, 22027, 22029, 22030, 22032, 22033, 22034, 22036, 22039, 22040, 22041, 22042, 22043, 22045, 22046, 22047, 22048, 22049, 22050, 22052, 22053, 22054, 22056, 22057, 22059, 22064, 22067,
Other devices under the same FDA submission
10 devices cleared under K093514, K111445 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Discovery 670 PRO H3100AHG E Medical Systems Israel Ltd. · System, tomography, computed, emission2 recalls0 reports
- Discovery 670 DRG E Medical Systems Israel Ltd. · System, x-ray, tomography, computed2 recalls0 reports
- NM 830G E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall1 report
- Discovery 630 H3100JYG E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall0 reports
- Discovery 670 ES H3100AGG E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall0 reports
- Discovery NMCT 670 ESG E Medical Systems Israel Ltd. · System, x-ray, tomography, computed1 recall0 reports
- Discovery NM 630G E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall0 reports
- Discovery 670 DRG E Medical Systems Israel Ltd. · System, tomography, computed, emission1 recall0 reports
- NMCT 870 DRG E Medical Systems Israel Ltd. · System, x-ray, tomography, computed1 recall0 reports
- D630 Upgrade H2401XFGE Hualun Medical Systems Co., Ltd · System, tomography, computed, emission1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page