TriVascular brand Fill Polymer Kit, Endovascular Graft… recallZ-0288-2013 · Trivascular, Inc · initiated Oct 19, 2012
Class II Terminated All recalls posted in November 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Trivascular, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Selected lots of the Fill Polymer Kit (European Model Number: TVFP14- B) in European distributor inventory only that potentially contain leaking stopcocks due to possible cracks in the stopcock component. TriVascular indicated that these Fill Polymer Kits are for use with the Ovation or Ovation Prime Abdominal Stent Graft Systems that may display leaks upon initiation of mix but prior to use in
FDA root cause category: Nonconforming Material/Component
What the company is doing
Trivascular sent an Urgent Field Safety Notice letter dated October 19, 2012 to all affect consignees. A verbal (telephone) communication was also initiated. The forms communication identified the affected product, problem and actions to be taken. Consignees were instructed to immediately quarantine and return affected products. A return response form was included for consignees to complete and returned via the e-mail address or fax number provided. Consignees were instructed to further notify its (sub-level) consignees, and have them return affected product directly to the recalling firm. For questions call 1-707-543-8732.
Recall details
- Recall number
- Z-0288-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 19, 2012
- Posted by the FDA
- Nov 13, 2012
- Recalling firm
- Trivascular, Inc
- Quantity
- 41 units
- Product code
- MIH
- FDA submissions
- P120006
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Belgium, Cyprus, Italy, Spain, Switzerland and Turkey. TriVascular stated that no purchased devices are in hospital inventory in Europe or the U.S.
Product description
TriVascular brand Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European Model Number: TVFP14-B, (in European distributor inventory only); PRODUCT Usage: The TriVascular Ovation Abdominal Stent Graft System is indicated in subjects diagnosed with an aneurysm in the abdominal aorta having vascular morphology suitable for endovascular repair, including: (1) Adequate iliac/femoral access compatible with vascular access techniques, devices, and/or accessories, (2) Non-aneurysmal proximal aortic neck: - with a length of at least 7 mm proximal to the aneurysm, - with an inner wall diameter of no less than 16 mm and no greater than 30 mm, and - with an aortic angle of d 60 degrees if proximal neck is e 10 mm and d 45 degrees if proximal neck is < 10 mm, (3) Non-aneurysmal distal iliac landing zone: - with a length of at least 10 mm, - with an inner wall diameter of no less than 8 mm and no greater than 20 mm
Code information (models, lots, serial numbers)
The following lot numbers are in distributor inventory in Europe: FF121211-04, FF010912-01, FF013112-01, FF020912-01, FF022212-03, FF022912-01, FF040412-01, FF051012-01
Other devices under the same FDA submission
42 devices cleared under P120006 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Ovation iX Abdominal Stent Graft System Aortic BodyTrivascular, Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment11 deaths105 reports
- Ovation iX Abdominal Stent Graft System Aortic BodyTrivascular, Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment4 deaths87 reports
- Ovation iX Abdominal Stent Graft System Aortic BodyTrivascular, Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths72 reports
- Alto™ Abdominal Stent Graft System TV-AB2980-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment1 death65 reports
- Alto™ Abdominal Stent Graft System TV-AB2680-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment5 deaths56 reports
- Alto™ Abdominal Stent Graft System TV-AB3480-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment6 deaths56 reports
- Ovation iX Abdominal Stent Graft System Aortic BodyTrivascular, Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths54 reports
- Alto™ Abdominal Stent Graft System TV-AB2380-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment4 deaths48 reports
- Alto™ Abdominal Stent Graft System TV-AB2080-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment20 reports
- Ovation iX Abdominal Stent Graft System Aortic BodyTrivascular, Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment20 reports
- Alto™ Abdominal Stent Graft System TV-AB2680-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment8 reports
- Alto™ Abdominal Stent Graft System TV-AB3480-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment1 death8 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page