SenTec V-Sign Membrane Changer (VS-MC), 9 pieces. The… recallZ-0433-2013 · Sentec AG · initiated Nov 5, 2012

Class II Terminated All recalls posted in November 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Sentec AG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A manufacturing defect was observed for the membrane changer insert that does not allow electrolyte to be present under the new membrane. The PCO2 performance of a sensor without electrolyte will be deteriorated.

FDA root cause category: Component design/selection

What the company is doing

SenTec notified the one US distributor via e-mail on November 5, 2012. The distributor was advised to stop distribution of affected lot numbers. The communication to the distributor described the issue, identified affected product, and stated that SenTec will replace current stock at the end-customer level. The end user can discard the affected products. The distributor is asked to respond back with an e-mail to SenTec that they took appropriate action.

Recall details

Recall number
Z-0433-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Nov 5, 2012
Posted by the FDA
Nov 22, 2012
Recalling firm
Sentec AG
Quantity
195 boxes (9 per box)
Product code
LKD
FDA submissions
K101690
Distribution
Distributed only in MO.

Product description

SenTec V-Sign Membrane Changer (VS-MC), 9 pieces. The SenTec V-Sign Membrane Changer is indicated for use with the V-Sign Sensor only. The V-sign Sensor is indicated for use with the SenTec Digital Monitor when continuous non-invasive monitoring of carbon dioxide tension, oxygen saturation, and pulse rate are required for adult and pediatric patients.

Code information (models, lots, serial numbers)

Lot numbers: 1CF5, 1D21, 1D60, 1DC6, 1DED, 1E0D

Other devices under the same FDA submission

6 devices cleared under K101690 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 63631).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page