Siemens Healthcare ADVIA Centaur Theophylline 2 (THE02) recallZ-0512-2013 · Siemens Healthcare Diagnostics, Inc · initiated Nov 14, 2012
Class II Terminated All recalls posted in December 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Operating range of the ADVIA Centaur systems is 18C to 30C (64.4F to 86.0F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.
FDA root cause category: Device Design
What the company is doing
Siemens Healthcare issued An Urgent Medical Device Correction Notice (for US customers) and an Urgent Field Safety Notice (for customers outside the US) dated November 5, 2012, to all affected customers. These notices inform customers who used the ADVIA Centaur BNP, BR, CA 19-9, Digitoxin, Folate, and Theophylline 2 assays to reduce their instrument operating temperature range to 20 -25 degrees C and for the DHEAS assay to reduce the instrument operating temperature range to 18 -26 degrees C when these assays are being used. These notices also direct customers to review the contents of the correction notice with their Laboratory Director or Medical Director. A confirmation fax-back form was included to ensure that all customers worldwide have been notified and have complied with the mandatory action. Customers with questions were instructed to contact their Siemens Technical Solutions Center or their local technical support representative. For questions regarding this recall call 508-668-5000.
Recall details
- Recall number
- Z-0512-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 14, 2012
- Posted by the FDA
- Dec 11, 2012
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Quantity
- 50 test=2397 and 250 test= 594
- Product code
- KLS
- FDA submissions
- K982680
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to the following countries: Argentina Australia Austria Azerbaijan Belgium Bosnia Herzeg. Brazil Canada Chile COLOMBIA Croatia Czech Republic Denmark Egypt Estonia Finland France Germany Greece Hong Kong Hungary Iceland India Indonesia Iran Ireland Italy Japan Kazakhstan Latvia Lithuania Malaysia Mexico Netherlands New Caledonia Norway Oman P.R. China Philippines Poland Portugal Qatar Rep. of Yemen Republic Korea Reunion Romania Russian Fed. Saudi Arabia Serbia Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey U.A.E. United Kingdom Uruguay Vietnam
Product description
Siemens Healthcare ADVIA Centaur Theophylline 2 (THE02) 250 Test 08838753 50 Test 03009619 Intended Use For in vitro diagnostic use in the quantitative determination of theophylline in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems.
Code information (models, lots, serial numbers)
All lot codes
Other devices under the same FDA submission
5 devices cleared under K982680 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Atellica IM BR 10995478Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR11 reports
- Atellica IM BR 10995477Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR8 reports
- ADVIA Centaur® BR +G 10333349Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR6 reports
- ADVIA Centaur® BR +G 10334837Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR4 reports
- ADVIA Centaur® BR 10340081Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 63682).
- Z-0507-2013 · TerminatedSiemens Healthcare BR Assay for CA 27.29 250 Test Kits 03896216 50 Test Kit 02419937 Ref Kit (US only) 10340081 Intended Use The ADVIA Centaur BR assay is an in vitro diagnostic …
- Z-0508-2013 · TerminatedSiemens Healthcare ADVIA Centaur CA 19-9 250 Test 10491244 50 Test 10491379 Intended Use The ADVIA Centaur CA 19-9 Assay is an in vitro immunoassay for the quantitative measureme…
- Z-0511-2013 · TerminatedSiemens Healthcare ADVIA Centaur Folate (FOL) 500 Test 06891541 100 Test 06367974 Ref Kit (US only) 09132781 Ref Kit (US only) 00203473 Intended Use For …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page