Bio-Modular Humeral Head 48mm x 24mm Ti 6AI4V/Ion Implanted… recallZ-0474-2013 · Biomet, Inc. · initiated Nov 1, 2012

Class II Terminated All recalls posted in December 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The items in this lot are missing the ion implantation feature. The surface of the head may be softer and more prone to scratching, which in tum could potentially cause higher wear of the poly bearing. Increased wear could potentially lead to a shorter implant life.

FDA root cause category: Process control

What the company is doing

Biomet, Inc. sent an URGENT MEDICAL DEVICE RECALL NOTICE dated November 1, 2012 to all affected consignees. The letter identified the affected product, problem and actions to be taken. The notification letter instructed the surgeon to montior the patient as their is a potential for increased wear which have an unknown effect on implant life. For questions call 574-372-1570.

Recall details

Recall number
Z-0474-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Nov 1, 2012
Posted by the FDA
Dec 5, 2012
Recalling firm
Biomet, Inc.
Quantity
4
Product code
KWT
FDA submissions
K030710
Distribution
USA Nationwide Distribution in the state of WI.

Product description

Bio-Modular Humeral Head 48mm x 24mm Ti 6AI4V/Ion Implanted (Orthopedic Implant) Product Usage: The OSS Knee Reamer Sleeves are primarily used to guide the bullet tip of cylindrical reamers during the tibial preparation in OSS procedures. They are reusable.

Code information (models, lots, serial numbers)

Catalog Number: 113740 and Lot Number Identification: 779270

Other devices under the same FDA submission

6 devices cleared under K030710 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.