Echo Bi-Metric Press-Fit Stems/Echo Bi-Metric 11mm x 135mm… recallZ-0442-2013 · Biomet, Inc. · initiated Nov 6, 2012
Class II Terminated All recalls posted in November 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
It was reported that the surgeon noticed that the Echo 11mm hip stem had an etching indicating it was 13mm stem.
FDA root cause category: Labeling Change Control
What the company is doing
The firm, Biomet, sent an "URGENT MEDICAL DEVICE RECALL NOTICE" dated November 8, 2012 to its customers. The notice describes the product, problem and actions to be taken. The customers were instructed to take the following actions: Immediately locate and remove the identified device(s) listed below from circulation. Carefully follow the instructions on the enclosed FAX Back Response Form. Fax a copy of the Response Form to 574-372-1683 prior to return of product. Use priority carrier for your shipment. Questions related to notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Recall details
- Recall number
- Z-0442-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 6, 2012
- Posted by the FDA
- Nov 27, 2012
- Recalling firm
- Biomet, Inc.
- Quantity
- 4
- Product code
- KWA
- FDA submissions
- K070274
- Distribution
- Nationwide distribution: USA including states of: AZ, MD, and KS.
Product description
Echo Bi-Metric Press-Fit Stems/Echo Bi-Metric 11mm x 135mm Collarless Porous Stem (Orthopedic Implant) Indications: 1) Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis, 2) rheumatoid arthritis, 3) correction offunctional deformity, 4) treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques, 5) revision of previously failed total arthoplasty.
Code information (models, lots, serial numbers)
Catalog Number: 192411 and Lot Number Identification: 244800
Other devices under the same FDA submission
38 devices cleared under K070274 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ECHO BI-METRIC HIP SYSTEM 192414Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 recall0 reports
- Echo® Bi-Metric® Hip System 192113Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish4 reports
- Echo® Bi-Metric® Hip System 192413Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED4 reports
- Echo® Bi-Metric® Hip System 192014Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal / polymer, non-porous, calcium phosphate3 reports
- ECHO BI-METRIC HIP SYSTEM 192109Biomet Orthopedics, LLC · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer2 reports
- Echo® Bi-Metric® Hip System 192009Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish2 reports
- Echo® Bi-Metric® Hip System 192010Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 reports
- Echo® Bi-Metric® Hip System 192011Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal / polymer, non-porous, calcium phosphate2 reports
- Echo® Bi-Metric® Hip System 192012Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal / polymer, non-porous, calcium phosphate2 reports
- Echo® Bi-Metric® Hip System 192112Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 reports
- Echo® Bi-Metric® Hip System 192409Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish2 reports
- Echo® Bi-Metric® Hip System 192410Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal / polymer, non-porous, calcium phosphate2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page