Biomet Offset Tibial Tray 2.5 mm Adaptor, REF 141490 recallZ-0533-2013 · Biomet, Inc. · initiated Nov 15, 2012
Class II Terminated All recalls posted in December 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Biomet investigation identified that the square end of the offset tibial tray adaptor locking insert is oversized and may not engage insert tool screwdriver. This could cause a possible delay in surgery greater than 30 minutes, potentially exposing the patient to the increased risks from being under anesthesia for a longer duration.
FDA root cause category: Employee error
What the company is doing
Biomet Recall sent an Urgent Medical Device Recall Notices letter dated November 15, 2012 to distributors, hospital operating room(OR)managers and Surgeons receiving the affected product. The letter identified the affected product, the reason for the recall and actions to be take, including directions to immediately locate and discontinue use of the affected product. Customers were asked to confirm receipt of the notice by calling 1-888-348-9500 extension 1570. Customers with affected product to be returned are asked to obtain RGA # by calling 574-372-6677 or email: [email protected]. A FAX Back Response Form is to be filled out and sent within 3 days of receipt to 574-372-1683. Return product to the Attention of Audrey Daenzer, Field Action Coordinator, Biomet, Attention Return Goods, Building B Recalls, 56 East Bell Drive, Warsaw, IN 46580. Questions related to the notice should be directed to 474-372-3983 or 574-371-3009, Monday- Friday 8 a.m. to 5 p.m.
Recall details
- Recall number
- Z-0533-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 15, 2012
- Posted by the FDA
- Dec 17, 2012
- Recalling firm
- Biomet, Inc.
- Quantity
- 200
- Product code
- MBV
- FDA submissions
- K010212
- Distribution
- Worldwide distribution - USA (nationwide) and the countries of Australia, Canada, Chile, Europe, Japan, Korea, Panama, Puerto Rico and Thailand.
Product description
Biomet Offset Tibial Tray 2.5 mm Adaptor, REF 141490 Sterile, QTY.1 Product Usage: Knee joint replacement prostheses intended for application with or without bone cement. Knee joint replacement components include femoral, tibial, and patellar components, intended for both primary and revision applications. Indications: 1) painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis when one or more compartments are involved, 2) correction of varus, valgus, or posttraumatic deformity, 3) correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure
Code information (models, lots, serial numbers)
Catalog Number: 141490 Lot 378330,378320,407230,485010,505140,485000,506070,533400,54048, 439700,797820
Other devices under the same FDA submission
19 devices cleared under K010212 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- BIOMET KNEE SYSTEM 141215Biomet Orthopedics, LLC · Prosthesis, knee, patello/femorotibial, semi-constrained, uhmwpe, pegged, cemented, polymer/metal/polymer1 recall3 reports
- Biomet® Knee System 141483Biomet Orthopedics, LLC · Prosthesis, knee, patello/femorotibial, semi-constrained, uhmwpe, pegged, cemented, polymer/metal/polymer2 reports
- Biomet® Knee System 141211Biomet Orthopedics, LLC · Prosthesis, knee, patello/femorotibial, semi-constrained, uhmwpe, pegged, cemented, polymer/metal/polymer2 reports
- Biomet® Knee System 141212Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER2 reports
- Biomet® Knee System 141213Biomet Orthopedics, LLC · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive2 reports
- BIOMET KNEE SYSTEM 141212Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 report
- Biomet® Knee System 141491Biomet Orthopedics, LLC · Prosthesis, knee, patello/femorotibial, semi-constrained, uhmwpe, pegged, cemented, polymer/metal/polymer1 report
- BIOMET KNEE SYSTEM 141216Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 report
- BIOMET KNEE SYSTEM 141213Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 report
- Biomet Knee System 141264Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Biomet Knee System 141263Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 report
- BIOMET KNEE SYSTEM 141481Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.