Alere Cholestech LDX Calibration Verification, 4 vial set… recallZ-0747-2013 · Alere San Diego, Inc. · initiated Oct 22, 2012
Class III Terminated All recalls posted in January 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Alere San Diego, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Out of range (increased) HDL Cholesterol results.
FDA root cause category: Component design/selection
What the company is doing
A recall letter was sent on 10/22/12 to all customers who purchased the Alere Cholestech LDX Mutianalyte Control, 2 vial set, Catalog Number 88769, Lot #C2510 & Lot #C2510A. The letter informed the customers that these lots of controls may yield results which are out of range high for the Level 2 control when tested with the Alere Cholestech LDX HDL test. Customers are instructed of the actions to be taken. On 11/16/12, the recall expanded to include the Alere Cholestech LDX Calibration Verification, 4 vial set, Catalog 88770, Lot#271097. The problem identified is that these lots of calibration verification materials may yield results which are out of range high when tested with the Alere Cholestech LDX HDL test. Customers are instruted of the actions to be taken. On 3/6/13, the recall expanded to include the Alere Cholestech LDX Calibration Verification, 4 Vial set, Catalog # 88770, Lot#298148 and 301886. The recall expanded to include the Alere Cholestech LDX Multianalyte Control, 2 vial Set, Catalog # 88769, Lots#C2572, C2572A, C2593 and C2593A. Customers are instructed to complete and fax the enclosed verification form within 10 days. Customers with questions are instructed to contact Alere San Diego, 9975 Summers Ridge Rd, San Diego, CA 92121, (877) 308-8289, fax (858) 805-8457, and [email protected].
Recall details
- Recall number
- Z-0747-2013
- Recall class
- Class III
- Status
- Terminated
- Initiated by the company
- Oct 22, 2012
- Posted by the FDA
- Jan 29, 2013
- Recalling firm
- Alere San Diego, Inc.
- Quantity
- 614 kits
- Product code
- JJY
- FDA submissions
- K102700
- Distribution
- Worldwide Distribution - USA (nationwide) Puerto Rico and South Korea.
Product description
Alere Cholestech LDX Calibration Verification, 4 vial set, Catalog Number: 88770 Alere Cholestech LDX Calibration Verification (hereafter referred to as CAL VERS) enables users to monitor the performance of total cholesterol (TC), high density lipoprotein cholesterol (HDL), triglycerides (TRG), and glucose (GLU) test procedures on the Alere Cholestech LDX System. Calibration verification is optional for CLIA waived systems, such as the Alere Cholestech LDX System, under CLIA regulations. However local or state regulations may require that calibration verification is run at regular intervals.
Code information (models, lots, serial numbers)
Lot#271097, 298148, and 301886
Other devices under the same FDA submission
1 device cleared under K102700 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Cholestech LDX™ 88773Alere San Diego, Inc. · Multi-Analyte Controls, All Kinds (Assayed)1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 63918).
- Z-0746-2013 · TerminatedAlere Cholestech LDX Multianalyte Control, 2 vial set, Catalog Number: 88769 Alere Cholestech LDX Multianalyte Control (hereafter referred to as Controls) enables users to monitor…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page