Spacelabs Smart Disclosure System, Model 92810, is an… recallZ-0694-2013 · Spacelabs Healthcare, Llc · initiated Dec 5, 2012
Class II Terminated All recalls posted in January 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Spacelabs Healthcare, Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Displaying or printing of an incorrect waveform for the time indicated in the printout or display in the Spacelabs Healthcare Intesys Clinical Suite (ICS) G2, Version 4.03, Clinical Access portion of Smart Disclosure, Model 92810.
FDA root cause category: Software design
What the company is doing
Spacelabs has sent the Urgent -Medical Device Correction Spacelabs Healthcare ICS G2 Version 4.03 Model 92810 Smart Closure letter, dated 19 December 2012, to all U.S. customers. The firm plans to send a customer letter (translated as necessary) via e-mail to all international subsidiaries and distributors of record on December 27, 2012. The letter is addressed to Hospital Administrator/Biomedical Manager/Office Manager. Users are advised on the following actions: 1) Advise your staff of this potential failure and advise them to always click the "Refresh" button when scrolling to new waveform data prior to printing or viewing waveform information or storing saved events. 2) We will contact you at the earliest possible date to schedule a convenient time for your hospital to upgrade your affected software at no cost. Customers with questions can contact Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support.
Recall details
- Recall number
- Z-0694-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 5, 2012
- Posted by the FDA
- Jan 17, 2013
- Recalling firm
- Spacelabs Healthcare, Llc
- Quantity
- 27 in US and 17 International
- Product code
- DQK
- FDA submissions
- K110779
- Distribution
- Worldwide Distribution-USA including Puerto Rico and the states of California, Florida, Iowa, Illinois, Louisiana, Michigan, Minnesota, North Carolina, South Carolina, Texas, and Washington, and the countries of Argentina, Mexico, France, and India.
Product description
Spacelabs Smart Disclosure System, Model 92810, is an integrated component of the Intesys Clinical Suite (ICS) G2, Version 4.03. The Smart Disclosure product collects patient waveforms, alarms, vital signs, and 12-lead reports. A 24-hour waveform acquisition is standard, with an upgrade option to 72 hours. The product provides basic trends in both tabular and graphical format, along with a retrospective ECG analysis. The analysis may be defined by patient, and the data may be presented in multiple formats, including a rhythm, a histogram, and a summary view. Standard reports are available from Smart Disclosure, including saved events, disclosure, trends, histogram, and summary. The Spacelabs Smart Disclosure System, Model 92810 is indicated for use in clinical situations where there is a need for review of physiological waveform information and alarm events up to 72 hours after the fact. Smart Disclosure is also indicated in those situations where a retrospective analysis of monitoring patients' ECG waveform data, that can be annotated and edited, is desired. The intended use of the Spacelabs Smart Disclosure is to interface with the Spacelabs monitoring network, providing the user with a means of recalling waveform information and performing retrospective analysis. The most recent 72 hours of monitored patient ECG waveform data can be analyzed, with each analysis limited to 24 hours or less.
Code information (models, lots, serial numbers)
ICS G2 Smart Disclosure Model 92810, Version 4.03. There are no serial numbers.
Other devices under the same FDA submission
1 device cleared under K110779 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Intesys® Clinical Suite Smart DisclosureSpacelabs Healthcare (Washington), Inc · COMPUTER, DIAGNOSTIC, PROGRAMMABLE1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page