Stryker Irrigator Reusable Tips recallZ-1220-2013 · Stryker Endoscopy · initiated Mar 27, 2013
Class II Terminated All recalls posted in May 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Endoscopy or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123C , instead of the correct 132-133C.
FDA root cause category: Error in labeling
What the company is doing
Stryker sent "Urgent Device Correction" letters dated March 27, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.
Recall details
- Recall number
- Z-1220-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Mar 27, 2013
- Posted by the FDA
- May 3, 2013
- Recalling firm
- Stryker Endoscopy
- Quantity
- 147,921 - all sizes
- Product code
- GCX
- FDA submissions
- K954726
- Distribution
- Worldwide Distribution: USA (nationwide) and countries of: Mexico, Portugal, Latin America, India, Greece, Taiwan, Singapore, New Zealand, Phillipines, Netherlands, France, Australia, Canada, Japan, Scandinavia, Iberia, Brasil, Romania, South Africa, Chile, Korea, China.
Product description
Stryker Irrigator Reusable Tips: 10mm X 32cm Pool Suction Tip, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.
Code information (models, lots, serial numbers)
Manufacturer Part Number: 250-070-404.
Other devices under the same FDA submission
1 device cleared under K954726 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- StrykeFlow 0250070505Stryker Corporation · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall3 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 64682).
- Z-1218-2013 · TerminatedStryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery…
- Z-1219-2013 · TerminatedStryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.
- Z-1221-2013 · TerminatedStryker Irrigator Reusable Tips: 5mm X 32cm Tip Hydrodissection Tip, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surg…
- Z-1222-2013 · TerminatedStryker Irrigator Reusable Tips: 5mm X 45cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.
- Z-1223-2013 · TerminatedStryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip , Non-Vented, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic sur…
- Z-1224-2013 · TerminatedStryker Irrigator Reusable Tips: 5mm X 24cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.
- Z-1225-2013 · TerminatedStryker Irrigator Reusable Tips: 5mm X 32cm Aspiration Needle Tip. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.
- Z-1226-2013 · TerminatedStryker Irrigator Reusable Tips: 2mm X 32cm Regular Tip with 1 Hole, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surg…
- Z-1227-2013 · TerminatedStryker Irrigator Reusable Tips: 2mm X 32cm Regular Tip with 2 Hole, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surg…
- Z-1228-2013 · TerminatedStryker Irrigator Reusable Tips: 2mm X 34cm Regular Tip with No Hole, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic sur…
- Z-1229-2013 · TerminatedStryker Irrigator Reusable Tips: 5mm X 28cm Regular Tip, vented, Non-Reflective Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic…
- Z-1230-2013 · TerminatedStryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip, vented, Non-Reflective Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic…
- Z-1231-2013 · TerminatedStryker Irrigator Reusable Tips: 10mm X 32cm Regular Tip, vented, Non-Reflective Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopi…
- Z-1232-2013 · TerminatedStryker Irrigator Reusable Tips: 5mm X 32cm Hydrodissection Tip, Non-Reflective Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic…
- Z-1233-2013 · TerminatedStryker Irrigator Reusable Tips: 5mm X 45cm Regular Tip, vented, Non-Reflective Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic…
- Z-1234-2013 · TerminatedStryker Irrigator Reusable Tips: Vented Tip, 5mm X 45cm Non-Reflective with 30 Downward Bend At Proximal End. Suction irrigator tips aid in cleaning and clearing the operative sit…
- Z-1235-2013 · TerminatedStryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip with No Holes, Non-Reflective Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparo…
- Z-1236-2013 · TerminatedStryker Irrigator Reusable Tips: 3mm X 32cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.
- Z-1237-2013 · TerminatedStryker Irrigator Reusable Tips: 3mm X 32cm Regular Tip, Non-Vented, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surg…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page