Synthes Radial Head Prosthesis: Trial Radial Head recallZ-1238-2013 · Synthes USA HQ, Inc. · initiated Feb 22, 2013
Class II Terminated All recalls posted in May 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes USA HQ, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Trial Head may come loose from the implant stem during manipulation of the arm during surgery. To date, no adverse events have been reported related to this issue and this recall is not being initiated as a result of adverse events.
FDA root cause category: Device Design
What the company is doing
Synthes sent an Urgent Notice: Medical Device Recall letter dated February 22, 2013, to all affected consignees. The letter requested consignees examine their inventory, remove them from use and return them to Synthes. Customers were asked to complete the attached Verification Section at the end of the letter indicating whether or not they have any of the affected product. For questions customers were instructed to call 610-719-5450 or email [email protected]. For questions regarding this recall call 610-719-5000.
Recall details
- Recall number
- Z-1238-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Feb 22, 2013
- Posted by the FDA
- May 3, 2013
- Recalling firm
- Synthes USA HQ, Inc.
- Quantity
- 417
- Product code
- KWI
- FDA submissions
- K112030
- Distribution
- Nationwide Distribution: USA including states of: IL, IN, MA, MI, MO, NJ, NY, OH, OR, PA, SC, TX, VA, WA and WI.
Product description
Synthes Radial Head Prosthesis: Trial Radial Head Elbow joint prosthesis: Synthes Radial Head Prosthesis System is a two-piece modular system comprised of titanium alloy stem and cobalt chrome components with an integral screw and side-loading application to allow in-situ assembly.
Code information (models, lots, serial numbers)
All Lots of Part Numbers: 03.402.018, 03.402.020, 03.402.022, 03.402.024, 03.402.026, 03.402.028, 03.402.218, 03.402.220, 03.402.222, 03.402.224, 03.402.226, 03.402.228, 03.402.418, 03.402.420, 03.402.422, 03.402.424, 03.402.426, 03.402.428, 03.402.618, 03.402.620, 03.402.622, 03.402.624, 03.402.626, and 03.402.628.
Other devices under the same FDA submission
21 devices cleared under K112030 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 04.402.008SSynthes (U.S.A.) LP · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER15 reports
- NA 09.402.022SSynthes (U.S.A.) LP · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER13 reports
- NA 04.402.007SSynthes (U.S.A.) LP · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER11 reports
- NA 04.402.027SSynthes (U.S.A.) LP · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER11 reports
- NA 09.402.024SSynthes (U.S.A.) LP · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER7 reports
- NA 04.402.009SSynthes (U.S.A.) LP · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER4 reports
- NA 04.402.006SSynthes (U.S.A.) LP · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER3 reports
- NA 09.402.018SSynthes (U.S.A.) LP · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER3 reports
- NA 09.402.020SSynthes (U.S.A.) LP · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER3 reports
- NA 09.402.026SSynthes (U.S.A.) LP · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER3 reports
- NA 09.402.222SSynthes (U.S.A.) LP · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER3 reports
- NA 09.402.622SSynthes (U.S.A.) LP · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page