CONFIDENCE SPINAL CEMENT SYSTEM 11CC PLUS KIT, Product… recallZ-2133-2013 · DePuy Spine, Inc. · initiated Apr 17, 2013
Class II Terminated All recalls posted in September 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DePuy Spine, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During injection of cement, the water in the hydraulic pump leaks past the piston within the pump body resulting in the loss of pressure and inability to continue to inject cement.
FDA root cause category: Process control
What the company is doing
DePut Synthes Spine sent an Urgent Voluntary Product Recall Notification letter dated April 15, 2013, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the con signee. Consignees were instructed to review their inventory and quarantine for return any products listed in the recall notice., complete the enclosed Business Reply Form, indicating level of inventory of the affected product and return the form to Depuy via fax to 888-943-4897or email to [email protected] and to return the product with the enclosed UPS label to Stericycle. For questions regarding this recall call 508-977-6606.
Recall details
- Recall number
- Z-2133-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 17, 2013
- Posted by the FDA
- Sep 3, 2013
- Recalling firm
- DePuy Spine, Inc.
- Quantity
- 3896 units (Overall quantity distributed appears greater than quantities manufactured due to the distribution pathways)
- Product code
- NDN
- FDA submissions
- K060300
- Distribution
- Worldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WV, WY; Worldwide: Algeria, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, France, Germany, Hong Kong, India, Israel, Italy, Latvia, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Poland, Portugal, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UAE, and UK
Product description
CONFIDENCE SPINAL CEMENT SYSTEM 11CC PLUS KIT, Product Code: 2839-10-000 The CONFIDENCE Spinal Cement System is intended for percutaneous delivery of CONFIDENCE Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures
Code information (models, lots, serial numbers)
HPBB0K, HPBB3C,HPCB7J, HPDBD6, HPBB0L,HPBB3D,HPCB7W,HPDBF0, HPBB0M,HPBB3F,HPDBBC,HPDBG4, HPBB0N,HPBBWH,HPDBBD,HPDBGB, HPBB0P,HPBBWJ,HPDBBK,HPDBGC,HPBB0T,HPBBWL,HPDBBL,HPDBGJ, HPBB19,HPBBZ8,HPDBBY,HPDBGM,HPBB27,HPBBZD,HPDBCD,HPDBGN, HPBB29,HPBBZN,HPDBCF, HPDBGY,HPBB2B,HPCB7B,HPDBD0,HPBB2C, HPCB7C,HPDBD3,HPBB2G,HPCB7H,HPDBD4
Other devices under the same FDA submission
3 devices cleared under K060300 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CONFIDENCE SPINAL CEMENT SYSTEM 283910000Medos International Sàrl · CEMENT, BONE, VERTEBROPLASTY5 deaths163 reports
- CONFIDENCE SPINAL CEMENT SYSTEM 283913000Medos International Sàrl · CEMENT, BONE, VERTEBROPLASTY8 deaths146 reports
- CONFIDENCE SPINAL CEMENT SYSTEM 283905000Medos International Sàrl · CEMENT, BONE, VERTEBROPLASTY6 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 64937).
- Z-2132-2013 · TerminatedCONFIDENCE SPINAL CEMENT SYSTEM 5CC KIT, Product Code: 2839-05-000 The CONFIDENCE Spinal Cement System is intended for percutaneous delivery of CONFIDENCE Spinal Cement, which is …
- Z-2134-2013 · TerminatedCONFIDENCE SPINAL CEMENT SYSTEM KIT, Product Code: 2839-13-000 The CONFIDENCE Spinal Cement System is intended for percutaneous delivery of CONFIDENCE Spinal Cement, which is in…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page