Four Lead TUR irrigation Set, for use with UROMATIC Plastic… recallZ-1682-2013 · Baxter Healthcare Corp. · initiated May 15, 2013
Class II Terminated All recalls posted in July 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A firm's internal investigation led to the discovery of pinholes in the packaging of the products. The pinholes are not likely to be detected by the end user prior to use and could compromise the sterile barrier properties of the packaging.
FDA root cause category: Packaging
What the company is doing
Baxter sent an Urgent Product Recall letter dated May 15, 2013 to all affected custorners. The letter identified the affected product, problem and actions to be taken. Customers were requested to remove all affected products and return for credit by contacting Baxter Healthcare Center for Service at 888-229-0001. Customers were instructed to complete the attached Customer Reply Form and return it to Baxter by either faxing to 224-270-5457 or scanning and e-mailing the reply form to [email protected]
Recall details
- Recall number
- Z-1682-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 15, 2013
- Posted by the FDA
- Jul 10, 2013
- Recalling firm
- Baxter Healthcare Corp.
- Quantity
- 356,728 units
- Product code
- LJH
- FDA submissions
- K960787
- Distribution
- US Nationwide Distribution
Product description
Four Lead TUR irrigation Set, for use with UROMATIC Plastic Containers, Product Code 2C4013 - Approximate Length 85" (2.2. m) Usage: Baxter irrigation sets are intended for the delivery of irrigating solutions from the fluid container to the irrigation site during bladder irrigation or endoscopic procedures including cytoscopy, transurethral resection and arthroscopic procedures. the sets are intended for use with Baxter irrigating solutions in Uromatic or Arthromatic plastic containers.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
UR175554 UR176743 UR182030 UR188144 UR192047 UR192096 UR192260 UR194951 UR202044 UR204032 UR211698 UR214312 UR216069 UR222471 UR225334 UR229294 UR233569 UR233700 UR238915 UR243022 UR246009 UR249383 UR255802 UR260273 UR262980 UR263004 UR265918 UR283580 UR287284 UR288746 UR298216 UR304246 UR307249 UR314328 UR318063 UR321000 UR325464 UR334151 UR339317 UR344234 UR352484 UR352567 UR360255 UR363432 UR366161 UR375857 UR383794 UR385534 UR07J19029 UR07K29059 UR07L21039 UR08B21053 UR08C25151 UR08E15157 UR08G22167 UR08I12065 UR08K13086 UR08L18158 UR09A21211 UR09D02157 UR09E14084 UR09G17125 UR09H26140 UR09J30040 UR09J30065 UR09J30107 UR09L01146 UR10C25033 UR10D09142 UR10E21037 UR10G28129 UR10I28034 UR10L03024 UR10L22024 UR11A14022 UR11B24086 UR11B25125 UR11C25032 UR11F01028 UR11G08088 UR11H08011 UR11H09019 UR11I08043 UR11I23026 UR12A25043 UR12E18502 UR12H03011 UR12H17045 UR12I05030 UR12I06053 UR12I21052 UR12J18040 UR12J26019 UR12L12023 UR13B08053 UR13B15041 UR13C08036
Other devices under the same FDA submission
3 devices cleared under K960787 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Arthromatic/Uromatic 2C4041Baxter International Inc. · System, irrigation, urological40 reports
- Arthromatic/Uromatic 2C4040Baxter International Inc. · ARTHROSCOPE15 reports
- Arthromatic/Uromatic 2C4005Baxter International Inc. · ARTHROSCOPE4 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 65183).
- Z-1683-2013 · TerminatedTwo Lead Arthroscopic Irrigation Set, for use with ARTHROMATIC Plastic Containers, Product Code 2C4030 - Approximate Length 98" (2.5. m) Usage: Baxter irrigation sets are intended…
- Z-1684-2013 · TerminatedFour Lead Arthroscopic Irrigation Set, for use with ARTHROMATIC Plastic Containers, Product Code 2C4031 -Approximate Length 99" (2.5. m) Usage: Baxter irrigation sets are intended…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.