Skip to content

Optovue iVue with Normative Database (NDB) with Software… recallZ-1441-2013 · Optovue, Inc. · initiated May 14, 2013

Class III Terminated All recalls posted in May 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Optovue, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Colors from the NDB comparison for Ganglion Cell Complex thickness in the singular OU report in iVue version 3.0 (US) and 3.1(international) are different from those in the individual iWellness ID and OS reports.

FDA root cause category: Software design

What the company is doing

Optovue sent a Field Correction Notice letter dated May 16, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.

Recall details

Recall number
Z-1441-2013
Recall class
Class III
Status
Terminated
Initiated by the company
May 14, 2013
Posted by the FDA
May 31, 2013
Recalling firm
Optovue, Inc.
Quantity
~329 users
Product code
HLI
FDA submissions
K121739
Distribution
Distributed Nationwide and in Canada.

Product description

Optovue iVue with Normative Database (NDB) with Software Version 3.0 and 3.1. Intended for in vivo imaging.

Code information (models, lots, serial numbers)

iVue software version 3.0 and 3.1 3.1 is international version, only distributed to Canada but not currently installed in any Canadian facilities.

Other devices under the same FDA submission

3 devices cleared under K121739 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page