Vermed, Inc Radiotransparent Defibrillation Electrodes recallZ-1435-2013 · Heart Sync, Inc · initiated Apr 26, 2013
Class II Terminated All recalls posted in May 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Heart Sync, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
On 4/18/13 HEARTSYNC became aware of a recall from their wire/connector sub-assembly supplier which is used in the C100-Physio Electrodes. There were reports of equipment damage due to misalignment from tooling installed in August 2012. The result of this damage could render the device incapable of delivering the appropriate energy or shock to the patient.
FDA root cause category: Process design
What the company is doing
HeartSync sent an URGENT MEDICAL DEVICE RECALL notification letter dated April 26, 2013, to all consignees of Heart Sync for C100-Physio Radiotransparent Defrillation Electrodes. On April 29, 2013, an URGENT MEDICAL DEVICE RECALL notification was sent to Vermed Inc for the D304 Defrillation Electrodes . The letters identified the product, the problem, and the action to be taken by the customer. Customers were instructed to return the affected product to the firm for full credit. Customers were asked to complete the form and send back via fax to 734-213-5640 or call 734-213-5530. For questions regarding this recall call 734-213-5530. .
Recall details
- Recall number
- Z-1435-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 26, 2013
- Posted by the FDA
- May 30, 2013
- Recalling firm
- Heart Sync, Inc
- Quantity
- 17180 units for all part numbers
- Product code
- LDF
- FDA submissions
- K080421
- Distribution
- Worldwide Distribution - USA including AL, GA, MI, MN, NC, ND, OH, PA, SD, VA, VT, OH, TX, MA, and Internationally to Korea.
Product description
Vermed, Inc Radiotransparent Defibrillation Electrodes Monophasic or biphasic compatible defibrillation/cardioversion/pacing/ECG Monitoring electrodes. Used in connection with a defibrillator to deliver an electrical current (shock) to the patient's heart during cardiac emergencies.
Code information (models, lots, serial numbers)
Part Number: D304 and lot numbers:Y100112-02, Y102212-03, Y121312-18, Y012413-02, Y022013-01, Y032113-01
Other devices under the same FDA submission
4 devices cleared under K080421 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Curaplex 16384Bound Tree Medical, LLC · Electrode, electrodcardiograph, multi-function4 reports
- Curaplex 16242Bound Tree Medical, LLC · Electrode, electrodcardiograph, multi-function1 report
- Curaplex 16250Bound Tree Medical, LLC · Electrode, electrodcardiograph, multi-function1 report
- Curaplex 16381Bound Tree Medical, LLC · Electrode, electrodcardiograph, multi-function1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 65085).
- Z-1434-2013 · TerminatedHeart Sync C100-Physio Radiotransparent Defibrillation Electrodes Monophasic or biphasic compatible pacing/ cardioversion Defibrillation /ECG Monitoring electrodes Used in connect…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page