Skip to content

R3 (TM) XLPE ACETABULAR LINER UHMWPE, 20o OVERHANG, 36 MM… recallZ-1928-2013 · Smith & Nephew Inc · initiated May 29, 2013

Class II Terminated All recalls posted in August 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smith & Nephew Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

One batch of 60 mm 20o R3 XLPE liner may be laser etched and packaged as a 62 mm 20o liner and one batch of 62 mm 20o R3 XLPE liner may be laser etched and packaged as a 60 mm 20o liner.

FDA root cause category: Packaging process control

What the company is doing

All affected Smith & Nephew Sales Representatives were notified of problem and the recall via email on 5/29. They were instructed to cease use and distribution of the product and quarantine for return. Surgeons were notified by letter beginning on 6/12. International consignees were notified on 05/29 via email.

Recall details

Recall number
Z-1928-2013
Recall class
Class II
Status
Terminated
Initiated by the company
May 29, 2013
Posted by the FDA
Aug 8, 2013
Recalling firm
Smith & Nephew Inc
Quantity
53 units
Product code
MBL
FDA submissions
K113848
Distribution
Worldwide Distribution-USA including the states of PA, WA, TX, FL, IA, MI, WI, TN, KY, IN, NY, MO and OR, and the countries of India, Canada, Chile, Venezuela, Dubai, South Africa, Australia, Poland, Romania, Belgium, Germany and United Kingdom.

Product description

R3 (TM) XLPE ACETABULAR LINER UHMWPE, 20o OVERHANG, 36 MM ID, 62 MM OD, QTY: (1), STERILE H2O2, REF 71335762. Orthopaedic.

Code information (models, lots, serial numbers)

Batch No. 12FM00603

Other devices under the same FDA submission

47 devices cleared under K113848 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 65370).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page