Skip to content

HemosIL RecombiPlasTin 2G, Part Number 0020002950 (8 ml… recallZ-1784-2013 · Instrumentation Laboratory Co. · initiated Apr 9, 2013

Class II Terminated All recalls posted in July 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Instrumentation Laboratory Co. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Instrumentation Laboratory is recalling HemosIL RecombiPlasTin 2G (RPT2G), Part No. 0020009580 (8 mL Size) due to being Out of Specification.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Between April 9, 2013 and April 11, 2013, Instrumentation Laboratory distributed "Important Product Correction" notification letters to their U.S. and foreign customers informing them that the on-board instrument stability claim has been reduced to 3 days for all current and future product lots of HemosIL RecombiPlasTin 2G. A second notification to non-respondents was initiated and completed on May 7, 2013. Technical questions were referred to 1-800-678-0710, Option # 2.

Recall details

Recall number
Z-1784-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Apr 9, 2013
Posted by the FDA
Jul 22, 2013
Recalling firm
Instrumentation Laboratory Co.
Quantity
Kits per product:1)11080, 2)11603, 3)11487, 4)11483, 5)11290, 6)11562, 7)11535, 8)11577, 9)11505, 10)11708,11)11579,12)11580, 13&14) not distributed
Product code
GJS
FDA submissions
K070005
Distribution
Worldwide Distribution- US (nationwide) including states of: MA, ME, CT, NJ, NY, PA, MD, VA, NC, SC, GA, FL, AL, TN, MS, KY, OH, IN, MI, IA, WI, MN, SD, MT, IL, MO, KS, NE, LA, AR, OK, TX, CO, ID, AZ, CA, OR, WA, AK, RI, VT and the countries of: Angola, Argentina, Australia, Austria, Belgium, Bermuda, Brazil, Bulgaria, Canada, Caribbean, Chile, Colombia, Costa Rica, Czech, Denmark, Dubai, Ecuador, Egypt, France, Germany, Gibraltar, Great Britain, Greece, Guatemala, Hungary, India, Indonesia, Iran, Israel, Italy, Jamaica, Jordan, Korea, Kosovo, Kuwait, Lebanon, Lithuania, Malaysia, Malta, Mexico, Morocco, Netherlands, Norway, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad, Turkey, Ukraine, Uruguay, Vietnam, and Japan.

Product description

HemosIL RecombiPlasTin 2G, Part Number 0020002950 (8 ml size). The product is used for the evaluation of the extrinsic coagulation pathway and the monitoring of Oral Anticoagulant Therapy.

Code information (models, lots, serial numbers)

Part Number applies to all: 0020002950 - all product lots.  Lot Numbers: 1) N0716706 Exp. Date 5/31/2013, 2) N0817321 Exp. Date 5/31/2013, 3) E0917396 Exp. Date 9/30/2013, 4)) N0917378 Exp. Date 9/30/2013, 5) N1118029 Exp. Date 10/31/2013, 6) N0128669 Exp. Date 1/31/2014, 7) N0329291 Exp. Date 2/28/2014, 8) N0429596 Exp. Date 3/31/2014, 9) N0520056 Exp. Date 5/31/2014,10) E0720822 Exp. Date 7/31/2014, 11) N0821174 Exp. Date 8/31/2014, 12) N091388 Exp. Date 9/30/2014, 13) N0921522 Exp. Date 9/30/2014, 14) N0132676 Exp. Date 1/31/2015 and 15) N1222567 Exp. Date 1/31/2015.

Other devices under the same FDA submission

1 device cleared under K070005 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page