BD MAX Instrument, catalog #441916 recallZ-1662-2013 · Becton Dickinson & Co. · initiated May 31, 2013

Class III Terminated All recalls posted in July 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton Dickinson & Co. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Electrical power module installed on some in vitro diagnostic instruments does not meet industry standards as labeled.

FDA root cause category: Component change control

What the company is doing

Becton Dickinson (BD) notified their customers in the US by phone on May 31, 2013 and by an Urgent Field Correction Action letter dated May 2013. Customers were advised of the affected product, problem and actions to be taken. The letter informed customers that an onsite visit by a BD Service representative is planned to perform a system check so to determine if the instrument is affected. BD's field service team plans to immediately replace the module on site if the affected part is found. For questions contact BD Technical Services Department at 1-800-638-8663.

Recall details

Recall number
Z-1662-2013
Recall class
Class III
Status
Terminated
Initiated by the company
May 31, 2013
Posted by the FDA
Jul 3, 2013
Recalling firm
Becton Dickinson & Co.
Quantity
71 instruments
Product code
NJR
FDA submissions
K111860
Distribution
Worldwide Distribution - USA Nationwide and in the countries of: Japan, Singapore, Hong Kong, China, Australia, Europe, and Chile.

Product description

BD MAX Instrument, catalog #441916 Product Usage: The BD MAX System, used in conjunction with the appropriate extraction kits and reagents is capable of automated amplification and detection of target nucleic acid sequences for use in clinical, industrial, and research settings.

Code information (models, lots, serial numbers)

Serial Numbers: CT0173 to CT0249, CT0253 to CT0260, CT0262, CT0263, CT0265, CT0269, and CT0272

Other devices under the same FDA submission

2 devices cleared under K111860 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.