AngioDynamics RITA Main Cable, 9 Pin to 14 Pin, REF… recallZ-1675-2013 · Angiodynamics, INC · initiated Jun 11, 2013
Class II Terminated All recalls posted in July 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Angiodynamics, INC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The RITA Main Cable product contains the incorrect Instructions for Use.
FDA root cause category: Packaging process control
What the company is doing
Consignees were notified via certified mail on 06/11/2013 with an "Urgent Medical Device Field Safety Correction" notification. This included the correct Instructions for Use and the recommended actions.
Recall details
- Recall number
- Z-1675-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 11, 2013
- Posted by the FDA
- Jul 10, 2013
- Recalling firm
- Angiodynamics, INC
- Quantity
- 13 units
- Product code
- GEI
- FDA submissions
- K040989
- Distribution
- Worldwide Distribution, including the US, Austria, China, Italy, Russia and the UK.
Product description
AngioDynamics RITA Main Cable, 9 Pin to 14 Pin, REF 700-101892. RITA Main Cables are indicated for use with the RITA RF Generators and RITA electrosurgical devices.
Code information (models, lots, serial numbers)
Catalog #700-101892, Lot number: 30.12
Other devices under the same FDA submission
3 devices cleared under K040989 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- AngioDynamics 700-101731MAngiodynamics, Inc. · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES3 reports
- AngioDynamics 700-102648Angiodynamics, Inc. · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES2 reports
- Intelliflow 700-102941Angiodynamics, Inc. · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page