MedStream Programmable Infusion Pump , Implantable for the… recallZ-1691-2013 · Codman & Shurtleff, Inc. · initiated Jun 7, 2013
Class I Terminated All recalls posted in July 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Codman & Shurtleff, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Miscalibrated Fill Level Sensor may affect dosing level
FDA root cause category: Device Design
What the company is doing
Codman issued a Medical Device Correction Notification to Medical Facilities dated June 7, 2013 via FedX to inform them of the potential problem and a Worksheet (instructions) to Identify Pumps with Miscalibrated FLS during each patient's next scheduled refill session or sooner if patient is symptomatic. If a pump is identified as having a miscalibrated FLS, notify Codman Neuro using the same worksheet. In addition a sample letter was provided to Physi cans to provide to patients on the system. Follow the recommendations for managing patients who are identified as having a pump with a miscalibrated FLS. For additional information,contact your Codman Neuro representative, or contact Codman Neuro Clinical Support at 1-800-660-2660. Please report any malfunction or adverse event related to this device to Codman Neuro at 1-866-491-0974, Option 2.
Recall details
- Recall number
- Z-1691-2013
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Jun 7, 2013
- Posted by the FDA
- Jul 16, 2013
- Recalling firm
- Codman & Shurtleff, Inc.
- Quantity
- 5 US and and 360 OUS
- Product code
- LKK
- FDA submissions
- P890055
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of FL, NC, NJ, and TX, and the countries of Austria, Belgium, Spain, Great Britain, Germany, Italy, Ireland, Luxemburg, Russia, Saudi Arabia, Portugal, Greece, Egypt, Switzerland, and Turkey.
Product description
MedStream Programmable Infusion Pump , Implantable for the intrathecal delivery of Baclofen or Morphine Product Code: 91-4201US (40ml pump) 91-4201 (40ml pump) Intended for the intrathecal delivery of Baclofen or Morphine.
Code information (models, lots, serial numbers)
All serial numbers
Other devices under the same FDA submission
6 devices cleared under P890055 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CODMAN AP-03000HCerenovus, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 deaths32 reports
- CODMAN AP-03000LCerenovus, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE17 reports
- NA IP-37957Medos International Sàrl · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE6 reports
- CODMAN AP-03000MCerenovus, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE3 reports
- MEDSTREAM 91-4201USCerenovus, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE3 reports
- CODMAN AP-03000ULCerenovus, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 65420).
- Z-1690-2013 · TerminatedMedStream Programmable Infusion Pump- Implantable for the intrathecal delivery of Baclofen or Morphine Product Code:91-4200 (20ml pump) US 91-4200. Intended for the intrathecal de…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗