Siemens Mammomat Inspiration - full-field digital… recallZ-1694-2013 · Siemens Medical Solutions USA, Inc · initiated Feb 28, 2013

Class II Terminated All recalls posted in July 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Firm became aware of an unintended behavior when using the Mammomat Inspiration. The possibility exists that a malfunction may occur during Stereo or Tomo acquisitions if the tomo clutch was incorrectly adjusted. Although the possibility is rare, the system could move while a patient's breast is compressed resulting in a possible bruise. Firm released Update Instructions XP031/12/S for this iss

FDA root cause category: Device Design

What the company is doing

Siemens issued an Update Instructions XP031/12/S by letter to the affected customers. For affected sites, a Siemens Service Engineer will inspect the unit and make any adjustments needed. Following the inspection and adjustments, a proper calibration will be completed. For questions call 610-448-3237 and 610-446-4634.

Recall details

Recall number
Z-1694-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Feb 28, 2013
Posted by the FDA
Jul 11, 2013
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
3
Product code
MUE
FDA submissions
K122286
Distribution
United States Nationwide Distribution in the states of ND, OH and SC.

Product description

Siemens Mammomat Inspiration - full-field digital mammography Product Usage: Mammography

Code information (models, lots, serial numbers)

Model number 10140000, serial numbers 3118, 3243 and 3309

Other devices under the same FDA submission

1 device cleared under K122286 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.