ADVIA Centaur HCY Diluent - 10mL Bottle Catalog number… recallZ-1960-2014 · Siemens Healthcare Diagnostics, Inc · initiated Jun 25, 2013
Class II Terminated All recalls posted in July 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
ADVIA Centaur Homocysteine dilution recovery on the diluent identified the percent recovery for the 1:10 patient sample dilutions lower than reported in the Instructions for Use (IFU).
FDA root cause category: Device Design
What the company is doing
Siemens Healthcare sent an Urgent Field Safety Notice that was sent to all affected Siemens Healthcare Diagnostics customers in the United States on June 25, 2013. The Urgent Field Safety Notice was sent out for distribution to all customers outside the United States on June 24, 2013. These notices inform the customer that the 1:10 diluted patient recovery is lower than reported in the Instructions for Use (IFU). Customers should discontinue use of both manual and on board 1:10 dilutions of samples with the ADVIA Centaur HCY assay. Customers can continue to use a manual or on board 1 :2 dilution for samples with results outside of the assay range. Siemens has confirmed the performance at the 1 :2 dilution level meets the IFU ranges. Please retain this letter with your laboratory records, and forward this letter to those who may have received this product. We apologize for the inconvenience this situation has caused. If you have any questions, please contact your Siemens Technical Solutions Center or your local Siemens technical support representative. For further questions please call (508) 668-5000.
Recall details
- Recall number
- Z-1960-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 25, 2013
- Posted by the FDA
- Jul 2, 2014
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Quantity
- 51 units
- Product code
- LPS
- FDA submissions
- K013406
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of : Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Ireland, Italy Kazakhstan, Latvia, Liechtenstein, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland ,Turkey, Unit.Arab Emir and Vatican
Product description
ADVIA Centaur HCY Diluent - 10mL Bottle Catalog number 09877493 (124533), SMN 10310370, and Ancillary Pack Catalog number 03302138 (124581), SMN 10378677. Used on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur CP platforms. For in vitro diagnostic use in the quantitative determination of total homocysteine (HCY) in serum or plasma.
Code information (models, lots, serial numbers)
Lot #/Exp. Date: 88D1524 01 Jun 2014 88D3303 26 Nov 2013
Other devices under the same FDA submission
2 devices cleared under K013406 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ADVIA Centaur® Systems, ACS:180® Systems HCY Calibrator 10310376Siemens Healthcare Diagnostics Inc. · CALIBRATOR, PRIMARY1 recall0 reports
- Atellica IM HCY Calibrator 10995498Siemens Healthcare Diagnostics Inc. · Calibrator, primary1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page