Mobiletta Mira recallZ-1875-2013 · Siemens Medical Solutions USA, Inc · initiated May 2, 2013

Class II Terminated All recalls posted in August 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Siemens issued a Field Safety Alert about the potential risk to users when operating the Mobilett Mira. When a Mobilett Mira is regularly exposed to extensive mechanical strain (driven over high thresholds for doors or elevators), it is possible that one of its main wheel shafts could break because of material fatigue, resulting in the chassis falling on the ground on the affected wheel side. This

FDA root cause category: Device Design

What the company is doing

Siemens sent a Field Safety Notice dated May 2. 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Customers were instructed to : 1) Never step or sit on the unit or otherwise put substantial weight on it. 2) Avoid moving the unit over high thresholds for doors or elevators where possible, or do so carefully and slowly. 3) Make sure that no person had his/her extremities under the chassis, also when the unit is not moving. Update instructions were also released to alert customers of the potential risk and to provide a hardware change which is intended to improve corrosion resistance. An acknowledgement of receipt form was also sent out for the field safety notice and requested to be returned to Siemens. For further questions please call (888) 826-9702.

Recall details

Recall number
Z-1875-2013
Recall class
Class II
Status
Terminated
Initiated by the company
May 2, 2013
Posted by the FDA
Aug 5, 2013
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
39
Product code
IZL
FDA submissions
K111912
Distribution
US Distribution including the states of IA, SD, NC, WI, OH, NE, TX, KS, ID, MI, GA, AR and IL.

Product description

Mobiletta Mira Mobile X-ray System

Code information (models, lots, serial numbers)

Model number 10273100

Other devices under the same FDA submission

1 device cleared under K111912 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.