Synthes Trauma Nail System. The devices are indication for… recallZ-2058-2013 · Synthes USA HQ, Inc. · initiated Nov 2, 2012
Class II Terminated All recalls posted in August 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes USA HQ, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Recall was initiated due to the possibility that the outer pouch was compromised. The product is packaged in two pouches; a sterile, internal pouch within an external pouch. It is the outer, external pouch that may be compromised.
FDA root cause category: Packaging change control
What the company is doing
The firm initiated their recall of this product on November 2, 2012 by sending an "URGENT NOTICE: MEDICAL DEVICE RECALL" letter to all consignees. The letter explained the problem and gave instructions for consignees to examine their inventory and urged consignees to return the identified devices. The letter stated If returning the product would adversely impact their ability to provide necessary medical care to patients, they should re-sterilize the product per Pre-Vacuum Sterilization instructions contained in the Instruction for Use.
Recall details
- Recall number
- Z-2058-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 2, 2012
- Posted by the FDA
- Aug 26, 2013
- Recalling firm
- Synthes USA HQ, Inc.
- Quantity
- 1,011,783
- Product code
- HSB
- FDA submissions
- K011857, K033071, K033618, K040336, K040762
- Distribution
- Nationwide distribution.
Product description
Synthes Trauma Nail System. The devices are indication for bone fixation.
Code information (models, lots, serial numbers)
Trauma Ex Nails, all lot numbers up to and including lot # 7072305 and Trochanteric Fixation Nails, all lot numbers up to and including lot # 7060897
Other devices under the same FDA submission
203 devices cleared under K011857, K033071, K033618, K040336, K040762 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 03.010.502Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES18 reports
- NA 04.003.028Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES12 reports
- NA 456.305Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES9 reports
- EXPert Nail 04.004.449SSynthes GmbH · NAIL, FIXATION, BONE8 reports
- EXPert Nail 04.004.549SSynthes GmbH · NAIL, FIXATION, BONE8 reports
- EXPert Nail 04.013.000Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES8 reports
- EXPert Nail 04.004.001Synthes GmbH · NAIL, FIXATION, BONE7 reports
- EXPert Nail 04.013.046Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES7 reports
- EXPert Nail 04.034.449SSynthes GmbH · NAIL, FIXATION, BONE7 reports
- EXPert Nail 04.034.452SSynthes GmbH · NAIL, FIXATION, BONE7 reports
- NA 456.303Synthes GmbH · Screw, fixation, bone7 reports
- NA 456.315SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES7 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page