Medtronic MiniMed Paradigm Reservoirs, Model No: MMT-326A… recallZ-1986-2013 · Medtronic MiniMed · initiated Jul 3, 2013
Class II Terminated All recalls posted in August 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic MiniMed or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic is recalling certain lots of Medtronic MiniMed Paradigm Reservoirs MMT-326A (1.8mL) and MMT-332A (3.0mL) used with Medtronic Paradigm insulin pumps because they may have increased risk for leaking. A leak in the reservoir may result in delivery of less insulin than intended. In addition, if there is a leaky reservoir and an insulin blockage occurs in the infusion set, the pump may not
FDA root cause category: Under Investigation by firm
What the company is doing
An Urgent Medical Device Recall notice was sent to customers on July 3, 2013, who purchased the Medtronic Paradigm Reservoirs MMT-326A and MMT-332A. The recall letter informed the customers of the problems and the actions to be taken. Medtronic instructed customers that they will replace all reservoirs from the affected lots at no cost. Customers are instructed to refer to www.medtronicdiabetes.com/support/product-updates for a list of frequently asked questions. Customers are instructed to report adverse reactions or quality problems experienced with the use of Medtronic Diabetes insulin pumps, infusion sets or reservoirs to FDA's MedWatch Adverse Event Reporting Program. Customers are instructed to call for replacements at (866) 222-2584 Monday through Friday (8am-6pm) central time. Customers with an urgent issue may use the 24 hour helpline.
Recall details
- Recall number
- Z-1986-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 3, 2013
- Posted by the FDA
- Aug 20, 2013
- Recalling firm
- Medtronic MiniMed
- Quantity
- 669,672 units
- Product code
- FRN
- FDA submissions
- K001828, K032005
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Australia, Bermuda, Brazil, Canada, Chile, Columbia, Ecuador, El Salvador, Germany, Japan, Mexico, Netherland, Paraguay, Puerto Rico, Singapore, Uruguay and Venezuela.
Product description
Medtronic MiniMed Paradigm Reservoirs, Model No: MMT-326A (1.8mL) and MMT-332A (3.0mL) The model MMT-326A and MMT-332A are syringe type insulin reservoirs intended for use with Medtronic Paradigm series insulin infusion pumps. The two reservoirs are identical except for barrel length and as a result, maximum insulin capacity. The MMT-326A holds a maximum of 1.8ml of insulin whereas the longer MMT-332A reservoir can hold up to 3.0ml of insulin. Both reservoirs are supplied sterile and are labeled for single use. Typically, the reservoir is replaced every tow or three days at the time of infusion set replacement. The Paradigm reservoirs have a shelf life of three years post sterilization. Product Quantity Distributed (Int) 669,672 units
Code information (models, lots, serial numbers)
MMT-326A Lot Numbers: H8437486 H8441420 H8442973 H8451531 H8486688 H8489386 H8485398 H8500423 H8500472 H8503728 H8512826 H8503372 H8510440 H8539013 H8627745 MMT-332A Lot Numbers: H8416432 H8420977 H8422490 H8424676 H8452933 H8455959 H8457716 H8459557 H8461538 H8463297 H8464121 H8467888 H8469703 H8471745 H8473271 H8476270 H8478398 H8473106 H8492449 H8491921 H8494645 H8496561 H8509305 H8512566 H8515317 H8517079 H8521052 H8541843 H8584244 H8603292 H8604958 H8635301
Other devices under the same FDA submission
6 devices cleared under K001828, K032005 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- MiniMed™ Reservoir MMT-332AMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control2 deaths3,013 reports
- MiniMed Reservoir MMT-326AMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control1 death762 reports
- MiniMed Reservoir MMT-332AMedtronic Minimed, Inc. · Pump, infusion342 reports
- MiniMed™ Reservoir MMT-332AMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control214 reports
- MiniMed Reservoir MMT-326AMedtronic Minimed, Inc. · Pump, infusion184 reports
- MiniMed™ Reservoir MMT-326AMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control30 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page