Edwards EndoReturn Arterial Cannula, model ER21B and ER23B recallZ-1972-2013 · Edwards Lifesciences, LLC · initiated Jul 16, 2013
Class II Terminated All recalls posted in August 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Edwards Lifesciences, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Edwards Lifesciences is recalling certain lots of the EndoReturn Arterial Cannula that were built with an incorrect component. This component can create a performance problem with the IntraClude IntraAortic Occlusion device.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Edwards sent an Urgent Field Safety Notice letter dated July 16, 2013 via Fed-Ex. The letter identified the affected product, problem and action to be taken. The letter included a Customer Acknowledgement Form to be completed and faxed to 800-422-9329 Attn: Recall Coordinator. If the customer does not send back the acknowledgement form, two additional contacts are made to notify the customer and request return of affected product. This is tracked by Edwards customer service and reported each week. Customer were requested to assess their inventory to assure that all affected product is identified and returned. Returned products will be destroyed. For questions call Edwards Customer Service at 800-424-3278.
Recall details
- Recall number
- Z-1972-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 16, 2013
- Posted by the FDA
- Aug 14, 2013
- Recalling firm
- Edwards Lifesciences, LLC
- Quantity
- 855
- Product code
- DWF
- FDA submissions
- K971291
- Distribution
- US Nationwide including Canada and VA sites.
Product description
Edwards EndoReturn Arterial Cannula, model ER21B and ER23B Product Usage: The ThruPort EndoReturn arterial cannula kit (with or without hemostasis valve), an introducer, a guidewire,and a connector hub. The ThruPort catheter introducer sheath kit includes a wire reinforced cannula with hemostasis valve, an introducer, and a guidewire. The cannulae have a wire reinforced section to provide kink resistance and flexibility, tapered tips to aid in insertion and advancement into the femoral artery, a non-reinforced clamp site, a 3/8 in. (9.5 mm) barbed connector, and a hemostasis valve (available on some versions) that allows passage of catheters, such as the EndoClamp aortic catheter. The introducers accept a .038 in. (0.97 mm) guidewire and are marked to simplify assembly and indicate alignment. On some versions, a connector hub secures and immobilizes the introducer within the cannula for easier, one person insertion of the cannula/introducer assembly.
Code information (models, lots, serial numbers)
Model ER21B Lots: 59496984, 59516933, 59516936, 59531385, 59531386 , 59543708 Model ER23B Lots: 59464327, 59484349, 59484350, 59484351, 59512146, 59513272, 59516940, 59534430, 59535739, 59541763
Other devices under the same FDA submission
1 device cleared under K971291 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ThruPort Systems EndoReturn Arterial Cannula Kit ER21BEdwards Lifesciences LLC · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page