GYRUS ACMI, REF 3740, Everest BiCOAG Bipolar Dissecting… recallZ-1983-2013 · Gyrus Medical, Inc · initiated Jul 24, 2013
Class II Terminated All recalls posted in August 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Gyrus Medical, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
On April 29, 2013 Olympus Surgical Technologies America R&D received results from an accelerated aging study conducted as part of a new market submission. The study found that one of 60 Dissector units that underwent 3 year accelerated aging failed the dye penetration test indicating that the seal on sterile packaging was compromised. The dissector products are labeled with a 3 year shelf life.
FDA root cause category: Packaging process control
What the company is doing
Olympus sent a "Medical Device Recall" letter dated July 24, 2013, to all affected customers. The letter identified the problem the product and the action needed to be taken by the customer.. Customers were advised to immediately cease any further use of the product and remove it from inventory and quarantine it until is shipped back to the company. To return the product, call 1-888-524-7266 Olympus customer service representative. Customers were requested to complete and return by fax the enclosed questionnaire to 484-896-7128. If the product was further distributed then they need to identify their customers and notify them at once of the product recall.
Recall details
- Recall number
- Z-1983-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 24, 2013
- Posted by the FDA
- Aug 19, 2013
- Recalling firm
- Gyrus Medical, Inc
- Quantity
- 1140
- Product code
- GEI
- FDA submissions
- K031080
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of AL, AZ, CA, FL, GA, IL, IN, KS,KY, MD, MA, MI, MN, MS, MO, NY, NC, OH, PA, SD, TN, TX, VA and WI., and the countries of BRAZIL, CHINA, AUSTRALIA, CHINA, GERMANY and INDIA.
Product description
GYRUS ACMI, REF 3740, Everest BiCOAG Bipolar Dissecting Forceps 5mm/45 cm, Sterile EO. GYRUS ACMI, REF 3740PK, or REF 94201PK, PKS LYONS, Dissecting Forceps 5mm/45cm, Sterile EO. The Dissectors are sterile, single use electrosurgical devices that are indicated for the electrosurgical coagulation, mechanical grasping, and dissection of tissue during the performance of laparoscopic and general surgical procedures.
Code information (models, lots, serial numbers)
REF 3740 150080AC 258624MC G166568CC 381809KD 408020AE JF480565 NY492562 REF 3740PK 128971JB 226730IC 226742IC 226743IC 431894CE 431895CE G128677JB REF 94201PK 102290EB 147428AC 167021CC 198596GC 204439HC 204451HC 249022KC 278142BD 320659DD 344712FD 363953ID 373761JD 389095MD 418231BE G386412KD G461037EE JF493195
Other devices under the same FDA submission
2 devices cleared under K031080 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- OLYMPUS 3700Gyrus ACMI, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES3 reports
- PK TECHNOLOGY 942005PKGyrus ACMI, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.