DRX Revolution Mobile X-Ray System, Model DRXR-1, Verona… recallZ-0090-2014 · Carestream Health, Inc. · initiated Jul 26, 2013
Class II Terminated All recalls posted in October 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Carestream Health, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During an audit, Carestream discovered a DRX-1 System that captured an image which displayed an artifact. This issue was discovered in manufacturing. It occurs when the DRX product console software is updated to v5.6b software and the detector is not calibrated before use.
FDA root cause category: Software design
What the company is doing
Carestream sent a Urgent-Medical Device Recall letter dated July 26, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Carestream Health Inc. recently discovered an image artifact issue due to a detector not being recalibrated with the 5.6B software upgrade. As our installation instructions did not require the user to recalibrate during an upgrade, there is a possibility your detector could do the same. When this occurs an image artifact could be evident on images captured in the Extended Exposure mode. These artifacts could potentially contribute to misdiagnosis or the necessity of patient re-exposure. We request that you take a moment to verify that your detector(s) are calibrated. At our next visit, Carestream will also verify that your detector was re-calibrated and, if not, re-calibrate the detector for you. Please note this only affects systems upgraded in the field. For questions or concerns, please contact the Carestream Customer Care Center at 1-800-328-2910, 8 am through 8 pm EST in the US.
Recall details
- Recall number
- Z-0090-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 26, 2013
- Posted by the FDA
- Oct 31, 2013
- Recalling firm
- Carestream Health, Inc.
- Quantity
- 23 units installed total (11 domestically & 12 internationally)
- Product code
- IZL
- FDA submissions
- K120062
- Distribution
- Worldwide Distribution - US (nationwide) including the states of New York, Massachusetts, Pennsylvania, Illinois, Texas, California and Louisiana., and the countries of Australia, Canada, Finland, Italy, Switzerland and United Kingdom.
Product description
DRX Revolution Mobile X-Ray System, Model DRXR-1, Verona Street Rochester, NY 14608 Catalog Numbers 1019397, 1023415, 1023423 --- Made in USA by Carestream Health, Inc. 150 The device is designed to perform radiographic x-ray examinations on all pediatric and adult patients, in all patient treatment areas.
Code information (models, lots, serial numbers)
Serial numbers: U.S. - 134,152,173,174,200,205,222,312,349,353,412, Worldwide Serial numbers: 469, 445, 306, 410, 453, 204, 483, 300, 171, 218
Other devices under the same FDA submission
1 device cleared under K120062 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- DRX-Revolution Mobile X-ray SystemCarestream Health, Inc. · System, x-ray, mobile2 recalls12 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.