Pipeline Embolization Device (PED) recallZ-2045-2013 · Micro Therapeutics Inc, Dba Ev3 Neurovascular · initiated Jul 19, 2013
Class II Terminated All recalls posted in August 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Micro Therapeutics Inc, Dba Ev3 Neurovascular or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Covidien Neurovascular has initiated the recall of several lots of Pipeline Embolization Device (PED) because the Instructions For Use (IFU) do not include the following contraindication: Patients in whom a pre-existing stent is in place in the parent artery at the target aneurysm location. Additionally, the designation of aspirin and clopidogrel was missing from the following contraindication "
FDA root cause category: Employee error
What the company is doing
Covidien sent an Customer Notification letter dated July 19, 2013, to all affected customers who received affected lots of Pipeline Embolization Device (PED). The customer notification letter informed the customers of the problems identified and the actions to be taken. Customer are instructed to reply and fax response form to Covidien Customer Service at (763) 398-7001. Customers are instructed that Customer Service will issue an RGA number for the return of the affected product back to Covidien. Customers are also instructed that upon receipt of a Returned Goods Authorization (RGA) number, to write the RGA number on the FEDEX slip and send the product back to: Covidien Attention: Product Recall 14005 13th Ave Plymouth, MN 55441 Prepaid and preaddressed FEDEX slips are enclosed for their convenience. Once Covidien receives the returned units, product replacement units will be shipped directly to customers. Local Covidien Neurovascular representative is also available to assist them in expediting the exchange of product. Customers with any questions, are instructed to call (949) 839-3700.
Recall details
- Recall number
- Z-2045-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 19, 2013
- Posted by the FDA
- Aug 23, 2013
- Recalling firm
- Micro Therapeutics Inc, Dba Ev3 Neurovascular
- Quantity
- 62 units
- Product code
- OUT
- FDA submissions
- P100018
- Distribution
- US Distribution including the states of AZ, KY, MO, NC, GA, WI, OR, FL, NY, CA, IL, TN, NV, TX, CO, MA, IA, UT and VA.
Product description
Pipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular treatment of adults (22 years of age and older) with large or giant wide-necked intracranial aneurysms in the internal carotid artery from the petrous to the superior hypophyseal segments.
Code information (models, lots, serial numbers)
Model/Lot Numbers: FA-71400-25 / 9747602 FA-77350-18 / 9749412 FA-77400-20 / 9759088 FA-71500-35 / 9747604 FA-77375-16 / 9744487 FA-77425-16 / 9752788 FA-77300-16 / 9757068 FA-77375-20 / 9757065 FA-77425-20 / 9744123 FA-77325-14 / 9759141 FA-77400-14 / 9757233 FA-77425-20 / 9752438 FA-77325-16 / 9753034 FA-77400-14 / 9768698 FA-77450-18 / 9753033 FA-77325-16 / 9757024 FA-77400-16 / 9743468 FA-77475-20 / 9752787 FA-77325-18 / 9759560 FA-77400-18 / 9751579 FA-77500-20 / 9750678 FA-77350-16 / 9752785 FA-77400-20 / 9755707
Other devices under the same FDA submission
292 devices cleared under P100018 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Pipeline TM Flex Embolization Device with Shield Technology TM PED2-300-12Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter2 recalls1 report
- PIPELINE PED-500-20Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall31 reports
- PIPELINE PED-500-35Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death29 reports
- PIPELINE PED-500-16Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall2 deaths27 reports
- PIPELINE PED-500-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death23 reports
- PIPELINE PED-500-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall21 reports
- PIPELINE PED-450-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall20 reports
- PIPELINE PED-475-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death19 reports
- PipelineTM Flex Embolization Device with Shield TechnologyTM PED2-500-12Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter67 reports
- PIPELINE PED-400-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall17 reports
- PIPELINE PED-475-14Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall16 reports
- PIPELINE PED-450-35Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death15 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page