MedStream Programmable Pump recallZ-2293-2013 · Codman & Shurtleff, Inc. · initiated Aug 13, 2013
Class I Terminated All recalls posted in September 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Codman & Shurtleff, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Drug flow rates exceeding programmed flow rates.
FDA root cause category: Use error
What the company is doing
Codman & Shurtleff, Inc. sent a Medical Device Field Safety Notice letter dated August 13, 2013 to US Consignees via FedEx overnight delivery and International Consignees were notified of the recall via email. The letter identifed the affected products, potential clinical impact, and updated IFU pump filling warning. The letter informed consignees that the Instructions for Use (IFU) and product training materials had been updated by adding a warning statement about potential over infusion if air gets inside the pump reservoir and additional clarifications to reinforce the proper filling technique. Consignees are asked to complete the Field Safety Notice Acknowledgement Form and return it to your local Codman Neuro Sales Representative or fax to: 508-977-6665. Consignees are asked to contact your Codman Neuro representative if they have questions, or contact Codman Neuro Clinical Support at 1-800-660-2660. Please report any malfunction or adverse event related to this device to 1-866-491-0974, Option 2.
Recall details
- Recall number
- Z-2293-2013
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Aug 13, 2013
- Posted by the FDA
- Sep 30, 2013
- Recalling firm
- Codman & Shurtleff, Inc.
- Quantity
- 2 units
- Product code
- LKK
- FDA submissions
- P890055
- Distribution
- Worldwide Distribution - USA Nationwide in the states of NJ, NC, TX, FL and in the countries of Austria, Belgium, Egypt, Germany, Greece, Ireland, Israel, Italy, Luxemborg, Portugal, Russia, Saudi Arabia, Slovenia, Spain, Sweden, Switzerland, Turkey, and UK.
Product description
MedStream Programmable Pump; 20ml - 91-4200US Product Usage: The Medstream programmable pump is intended for the intrathecal delivery of Baclofen in the US market and for the intrathecal delivery of Morphine or Baclofen in the EMEA market. Refill Kits- are intended for use in the filling and re-filling of the Medstream pump reservoir.
Code information (models, lots, serial numbers)
Lot/Serial Number US: NMBJ7K, NMBJ5C
Other devices under the same FDA submission
6 devices cleared under P890055 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CODMAN AP-03000HCerenovus, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 deaths32 reports
- CODMAN AP-03000LCerenovus, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE17 reports
- NA IP-37957Medos International Sàrl · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE6 reports
- CODMAN AP-03000MCerenovus, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE3 reports
- MEDSTREAM 91-4201USCerenovus, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE3 reports
- CODMAN AP-03000ULCerenovus, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 65821).
- Z-2294-2013 · TerminatedMedStream Programmable Pump; 40ml - 91-4201US Usage: The Medstream programmable pump is intended for the intrathecal delivery of Baclofen in the US market and for the intrathecal …
- Z-2295-2013 · TerminatedMedStream Refill Kit, U.S: 91-4287 Product Usage: The Medstream programmable pump is intended for the intrathecal delivery of Baclofen in the US market and for the intrathecal del…
- Z-2296-2013 · TerminatedMedStream Refill Kit (1 unit), EMEA: 91-4289 Product Usage: The Medstream programmable pump is intended for the intrathecal delivery of Baclofen in the US market and for the intra…
- Z-2297-2013 · TerminatedMedStream Refill Kit (6 units), EMEA: 91-4290 Product Usage: The Medstream programmable pump is intended for the intrathecal delivery of Baclofen in the US market and for the intr…
- Z-2298-2013 · TerminatedMedStream Refill Kit, U.S (6 units): 91-4288 Product Usage: The Medstream programmable pump is intended for the intrathecal delivery of Baclofen in the US market and for the intra…
- Z-2299-2013 · TerminatedMedStream Programmable Pump; 40ml Pump EMEA 91-4201 Product Usage: The Medstream programmable pump is intended for the intrathecal delivery of Baclofen in the US market and for th…
- Z-2300-2013 · TerminatedMedStream Programmable Pump; 20ml Pump EMEA 91-4200 Product Usage: The Medstream programmable pump is intended for the intrathecal delivery of Baclofen in the US market and for th…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page