ADVIA 120 Hematology Analyzer, Catalog Numbers SMN 10360958… recallZ-2114-2013 · Siemens Healthcare Diagnostics · initiated Jun 26, 2013
Class II Terminated All recalls posted in August 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The ADVIA 120 optics cover or hood is located on the very top of the ADVIA 120 analyzer. When the optics cover is raised, it is supported by two gas struts (SMN 10309266) attached to the cover on the left and right sides. Over time, it is possible that these struts may lose their effectiveness and fail to support the optics cover in the open position. This could cause the cover to fall during maintenance procedures.
FDA root cause category: Equipment maintenance
What the company is doing
Siemens sent an Urgent Device Correction Letter dated June 26, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed of the ADVIA 120 system displays a problem when raising and supporting the optics cover, immediately contact your service provider to arrange to have the cover struts serviced. If the optics cover will not stay in the up position, as required, do not perform maintenance procedures on the optics area until the hood struts have been replaced. Siemens field support will be proactively replacing the optics cover hood struts on a regular preventive maintenance cycle based on the age of the system. Please forward this notification to whomever you may have distributed this product We apologize for the inconvenience that this situation has caused. Thank you for your patience and continued support. This response form is to confirm receipt of the enclosed Siemens Healthcare Diagnostics Urgent Device Correction dated June 2013 regarding Optics Cover Struts (SMN 10309266). Please read each question and indicate the appropriate answer. Fax this completed form to Siemens Healthcare diagnostics at the fax number indicated at the bottom of this page. PLEASE FAX THIS COMPLETED FORM TO THE TECHNICAL SOLUTIONS CENTER AT (302) 631-7597.
Recall details
- Recall number
- Z-2114-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 26, 2013
- Posted by the FDA
- Aug 28, 2013
- Recalling firm
- Siemens Healthcare Diagnostics
- Quantity
- Domestic: 950 units; Foreign: 3369 units
- Product code
- GKL
- FDA submissions
- K971998
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Austria, Australia, Belgium, Bosnia - Herzegovina, China, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hungary, Iceland, India, Indonesia, Ireland, Iraq, Israel, Italy, Japan, Kenya, Republic of Korea, Kuwait, Lebanon, Lithuania, Luxembourg, Martinique, Malaysia, Netherlands, Norway, New Zealand, Philippines, Poland, Portugal, Romania, Russian Republic, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom and Vietnam.
Product description
ADVIA 120 Hematology Analyzer, Catalog Numbers SMN 10360958 and 10360959 . The ADVIA 120 is a hematology system that utilizes the principles of flow cytometry in order to provide complete blood counts.
Code information (models, lots, serial numbers)
All serial numbers
Other devices under the same FDA submission
1 device cleared under K971998 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ADVIA® 120 Hematology System 10360958Siemens Healthcare Diagnostics Inc. · Counter, differential cell1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.