Skip to content

Synthes Midfoot Fusion Bolt 6.5 mm. It is indicated for… recallZ-0014-2014 · Synthes USA HQ, Inc. · initiated Mar 8, 2013

Class II Terminated All recalls posted in October 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Synthes USA HQ, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A post market safety review following several customer complaints indicated additional clarification was necessary in the Midfoot Fusion Bolt Technique Guide to reduce potential for bolt back-out or migration by emphasizing the need for supplemental fixation.

FDA root cause category: Labeling design

What the company is doing

An "Urgent Notice: Medical Device Recall" letter dated March 8, 2013 was sent to customers asking them to check their inventory, remove all affected product technical guides from use, and return them to Synthes along with the Verification form. Customer questions were directed to (610) 719-5450.

Recall details

Recall number
Z-0014-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 8, 2013
Posted by the FDA
Oct 17, 2013
Recalling firm
Synthes USA HQ, Inc.
Quantity
62
Product code
HWC
FDA submissions
K081071
Distribution
Nationwide Distribution, including the states of MI, PA, SC, IA, MA, CA, WI, IL, and KS.

Product description

Synthes Midfoot Fusion Bolt 6.5 mm. It is indicated for fracture, fixation, osteotomies, nonunions, and fusions of large bones in the foot and ankle.

Code information (models, lots, serial numbers)

Part Nos.: 04.111.210, 04.111.220, 04.111.230. with Lot Nos.; 2294209 through 7587547.

Other devices under the same FDA submission

1 device cleared under K081071 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page