Edwards Lifesciences Fem-Flex Femoral Arterial Cannula with… recallZ-0134-2014 · Edwards Lifesciences, LLC · initiated Aug 26, 2013
Class II Terminated All recalls posted in November 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Edwards Lifesciences, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Edwards Lifesciences is recalling two lots of Duraflo Cannulas due to reported complaints involving cracks in the t-connector of the devices which were confirmed during product evaluation.
FDA root cause category: Process control
What the company is doing
Edwards sent an Urgent Field Safety Notice Product Recall letter dated August 26, 2013 to all affected customer, via Fed-Ex. The product is sold directly to hospitals and distributors. The distributors involved will communicate with their customers directly. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review their entire inventory, quarantine and return the affected products. An acknowledgement form is included for completion to assist in the assessment of the inventory. For questions contact Edwards Customer Service at 800-424-3278.
Recall details
- Recall number
- Z-0134-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Aug 26, 2013
- Posted by the FDA
- Nov 1, 2013
- Recalling firm
- Edwards Lifesciences, LLC
- Quantity
- 80 units
- Product code
- DWF
- FDA submissions
- K123370
- Distribution
- Nationwide Distribution.
Product description
Edwards Lifesciences Fem-Flex Femoral Arterial Cannula with Duraflo Coating, DIITFA02225, Sterile, RX only For Single Use Only, Product Usage: Edwards Lifesciences femoral access cannulae are intended for use in situations which require rapid femoral venous and arterial access for short-term (< 6 hours) cardiopulmonary bypass. Vessel access (venous or arterial) is left to the discretion of the physician. Femoral access cannulae may be used in pediatric populations or adult populations based on flow rate requirements and individual patient anatomy. Please consult labeling to determine pressure drop related to flow rates. Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.
Code information (models, lots, serial numbers)
Lot number 59388992 expiry 1/1/16
Other devices under the same FDA submission
7 devices cleared under K123370 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Edwards Lifesciences EZ Glide Aortic Cannula EZF21AEdwards Lifesciences LLC · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS8 reports
- Edwards Lifesciences EZ Glide Aortic Cannula EZC21AEdwards Lifesciences LLC · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS4 reports
- Edwards Lifesciences EZ Glide Aortic Cannula EZS21TAEdwards Lifesciences LLC · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS3 reports
- Edwards Lifesciences EZ Glide Aortic Cannula EZC24AEdwards Lifesciences LLC · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS3 reports
- Edwards Lifesciences EZ Glide Aortic Cannula EZC24TAEdwards Lifesciences LLC · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS3 reports
- Edwards Lifesciences EZ Glide Aortic Cannula EZC21TAEdwards Lifesciences LLC · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS1 report
- Edwards Lifesciences EZ Glide Aortic Cannula EZF24AEdwards Lifesciences LLC · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page