Brand Names: Integra Camino 110-4 Intracranial Pressure… recallZ-0117-2014 · Integra Limited · initiated Sep 10, 2013
Class II Terminated All recalls posted in October 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Integra Limited or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Integra LifeSciences has identified through an internal evaluation that in order to comply with US FDA requirements, certain Camino Intracranial Pressure Monitoring Kits require additional labelling to indicate they are UNSAFE in an MR environment. Only the following Camino Intracranial Pressure Monitoring Kits are affected: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC.
FDA root cause category: Under Investigation by firm
What the company is doing
Integra sent an Urgent Medical Device Correction Field Safety Notice dted September 10, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review their inventory and determine if they have any of the affected product. If so, customers were asked to ensure that the information contained in the letter be distributed to the appropriate personnel. Complete the attached form return it by email or fax as indicated on the form. Customers with questions were instructed to contact Customer Service at 1-855-532-1723. For questions regarding this recall call 1-800-997-4868, outside of the US 888-980-7742.
Recall details
- Recall number
- Z-0117-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 10, 2013
- Posted by the FDA
- Oct 30, 2013
- Recalling firm
- Integra Limited
- Quantity
- 79,887
- Product code
- GWM
- FDA submissions
- K102875
- Distribution
- Nationwide Distribution
Product description
Brand Names: Integra Camino 110-4 Intracranial Pressure (ICP) Monitoring Catheters. Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively. Common Name: Intracranial Pressure and Pressure-Temperature Monitoring Kits Classification Name: Device, Monitoring, Intracranial Pressure 510(k) Number: K102875 The Camino 110-4 Intracranial Pressure monitoring catheters are sterile transducer-tipped pressure monitoring catheters that are used by neurosurgeons to rapidly determine and continuously monitor intracranial pressure or pressure/temperature. The Camino 110-4B catheter is used for Intracranial Pressure Monitoring in the parenchyma or subarachnoid space. The Camino 110-4G catheter is used for Post Craniotomy Subdural Pressure Monitoring. The Camino 110-4H catheter is intended for Intracranial Pressure Monitoring and Cerebrospinal fluid drainage. The Camino 110-4L catheter is intended for use with the Licox Brain Oxygen Monitoring System to measure intracranial pressure in the parenchyma. The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively.
Code information (models, lots, serial numbers)
Show the full code information (10k characters)
Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. Lot Numbers: Catalog Number 110-4B: 3050RX195484, 3050RX198473, 305000205426, 3050RX197641, 3050RX201421, 305000205648, 305000205652, 3050RX199149, 3050RX201407, 3050RX199850, 305000205976, 3050RY189979, 3050RX200862, 305000206178, 3050RY193648, 3050RX200486, 3050RX181026, 3050RX185115, 3050RY195484, 305000207061, 305000207062, 3050RY198473, 305000207330, 305000207334, 3050RY197641, 3050RY200862, 305000207580, 305000207892, 3050RY200486, 3050RY181026, 3050RY185115, 3050RX207330, 3050RX207334, 3050RX207580, 305000208008, 305000208510, 3050RY207892, 305000210060, 3050RX208008, 3050RX208510, 3050RX205976, 3050RX206178, 305000210563, 305000211723, 3050RX210060, 305000212616, 305000213007, 305000213191, 30500X210563, 3050RX207061, 305000214590, 3050RX205426, 30500X215502, 3050RX207062, 305000215531, 30500X216410, 3050RY199850, 305000216132, 30500X216627, 30500X217119, 3050RY217112, 305000216718, 3050RX217751, 3050RX217840, 3050RY217837, 305000217858, 30500X218369, 30500X218391, 30500X219054, 3050RY206178, 305000220708, 305000218578, 30500X220430, 305000219261, 30500X221221, 305000221226, 305000219989, 30500X222569, 305000221953, 305000221955, 30500X223153, 30500X223709, 305000223275, 305000223280, 305000222640, 30500X224142, 305000223867, 305000223866, 30500X224692, 305000224352, 305000224353, 30500X225386, 305000225045, 305000225044, 30500X225992, 305000225632, 305000225633, 30500X226540, 30500X227059, 305000226844, 305000226846, 305000226127, 30500X227564, 305000226836, 30500X227820, 305000227127, 305000227128, 30500X228194, 305000227790, 305000227789, 30500X228787, 305000228522, 305000228521, 305000228971, 305000228972, 30500X229560, 305000229328, 305000229329, 30500X229766, 30500X230087, 305000229825, 305000229826, 30500X230565, 305000230303, 305000230302, 30500X231053, 305000230780, 305000230781, 30500X231786, 305000231461, 305000231462, 30500X232331, 305000232098, 305000232099, 30500X232960, 305000232605, 305000232604, 305000233245, 305000233244, 305000233243, 305000233242, 305000233241, 305000233240, 30500X233415, 30500X234183, 305000233607, 305000233608, 30500X234806, 305000234303, 305000234304, 305000235022, 305000235017, 30500X235732, 305000235598, 30500X236328, 305000236105, 305000236104, 305000236655, 305000236653, 30500Y237087, 30500X237090, 305000237232, 30500X237761, 305000237824, 305000237825, 30500X238273, 305000238263, 305000238264, 30500X239095, 305000238818, 305000238819, 30500X239497, 30500X239748, 30500X239958, 30500X240334, 30500X240668, 30500X241368, 305000241516, 305000242236, 305000242792, 305000242795, 305000242796, 305000242794, 305000242793, 305000242791, 3050RY243869, 305000243251, 305000243882, 305000243881, 305000243888, 305000243889, 305000243883, 30500X244281, 30500X244880, 305000244513,30500X245306, 305000244481, 305000244480, 305000244934, 30500X245937, 30500X246067, 305000245408, 30500X246872, 305000245835, 30500X246951, 30500X247608, 305000246460, 30500X248284, 305000247007, 3050RX227127, 3050RX234303, 305000247717, 305000249072, 3050RX242794, 3050RX242793, 3050RX237825, 3050RX241516, 30500X249488, 305000248479, 30500X250423, 305000249743, 30500X250964, 305000250456, 30500X251060, 305000251055, 30500X251560, 30500X252133, 3050RY252136, 305000251453, 30500X252771, 305000252058, 3050RX244513, 305000252646, 305000253785, 30500X254929, 3050RY255145, 30500X255141, 305000254589, 305000255549, 305000255551, 305000255038, 30500X256614, 305000256192, 305000257281, 305000256709, 305000257280, 305000258048, 30500X259385, 305000258379, 305000258921, 305000258922, 30500X259557, 305000259431, 305000259432, 30500X260159, 305000259966, 305000259960, 305000260514, 305000260527, 30500Y261748, 3050RX253785, 305000261325, 305000261326, 305000261737, 305000261736, 3050RX256192, 30500X262441, 305000262328, 305000262327, 305000262978, 305000262977, 3050RY263508, 30500X263559, 305000263504, 305000264073, 305000264074, 305000264625, 305000264624, 3050RY265348, 30500X265422, 30500X265976, 3050RY265975, 305000265530, 305000265527, 30500X266142, 305000266134, 305000266135, 30500X266789, 305000267193, 305000267194, 3050RX261325, 30500X267468, 305000267421, 305000267422, 30500X267990, 305000267673, 305000267674, 305000268294, 305000268293, 305000266784, 305000268914, 305000268915, 3050RX261736, 3050RY269955, 30500X269963, 305000269220, 305000269221, 305000270081, 3050RY271045, 30500X271047, 305000271042, 305000271716, 305000271715, 30500X272367, 305000272273, 305000272274, 305000272852, 305000272851, 30500X274204, 305000273466, 305000273463, 305000274200, 305000274201, 30500X275297, 30500X275393, 305000274816, 30500X276594, 3050RX268914, 305000277761, 3050RY282196, 305000282304, 30500X282815, 3050RY283371, 3050RX283315, 3050RX283314 Catalog Number 110-4G: 3050RX195650, 3050RX198690, 3050RY193969, 3050RY197837, 3050RX199151, 3050RX199942, 3050RX201552, 3050RY185259, 3050RX200538, 305000206177, 3050RX200989, 305000206556, 305000206557, 3050RX200990, 3050RY199151, 3050RY199942, 305000207465, 3050RY200538, 3050RY200989, 305000208007, 3050RY200990, 3050RX207465, 3050RX206557, 3050RX206556, 305000209417, 305000209105, 3050RX208007, 305000210891, 305000211419, 3050RX206177, 30500X209105, 3050RX209417, 305000212298, 30500X210891, 305000213240, 305000214103, 30500X215501, 30500X215577, 3050RX202201, 305000215628, 3050RY195650, 3050RY198690, 30500X216621, 305000216246,3050RY217113, 30500X217118, 305000217140, 30500X217835, 3050RY217839, 305000217464, 30500X218393, 305000218009, 30500X219053, 305000218225, 305000219118, 30500X220001, 30500X220429, 305000219990, 30500X221225, 305000221227, 30500X222567, 305000222642, 305000223279, 30500X223707, 305000223868, 30500X224685, 305000224354, 30500X225383, 305000225046, 305000225634, 30500X226000, 30500X226541, 305000226731, 305000226869, 305000226867, 30500X227567, 305000227129, 30500X228784, 305000228523, 305000228973, 30500X229561, 305000229330, 30500X230084, 305000229827, 305000230304, 30500X231054,305000230782, 30500X231787, 305000231463, 30500X232333, 305000232101, 30500X232776, 305000232606, 30500X233419, 305000233246, 305000233609, 30500X234690, 30500X234807, 305000234306, 305000235566, 30500X236329, 30500Y237088, 30500X237767, 305000237233, 305000238820, 30500X239496, 305000239476, 30500X240336, 30500X241357, 305000241518, 30500X242172, 305000242249, 305000243252, 305000244364, 30500X245304, 305000244935, 305000245409, 305000245838, 30500X246522, 305000246308, 30500X246873, 30500X247609, 30500X248281, 305000249073, 3050RX241518, 3050RX233246, 3050RX242249, 3050RX228973, 30500X251061, 305000251056, 30500X251563, 3050RY252137, 305000252059, 305000252648, 305000254591, 30500X254931, 30500X255143, 305000255039, 305000255548, 305000256193, 305000256737, 305000257244, 305000257246, 305000257921, 305000257920, 305000258303, 305000258290, 30500X259383, 305000258793, 305000259573, 30500X260163, 305000259958, 305000259959, 3050RX254591, 3050RX255039, 305000260513, 305000260512, 30500Y261771, 3050RX257244, 3050RX257920, 305000261159, 305000261160, 3050RY262444, 30500X262443, 305000261734, 305000261732, 305000262325, 305000262326, 3050RX258303, 305000262976, 305000262975, 3050RY263510, 30500X263560, 305000263507, 305000263506, 305000264075, 305000264076, 305000264627, 305000264628, 30500X265424, 305000265525, 305000265524, 30500X265978, 30500X266143, 305000266063, 305000266064, 30500X266787, 305000266638, 305000266637, 305000267191, 30500X267470, 305000267423, 305000267675, 3050RY269957, 30500X274206 Catalog Number 110-4H: 3050RX198281, 3050RY194144, 3050RY198281, 3050RY200861, 30500X223154, 305000220948, 305000223622, 30500X225385, 30500X233417, 305000231714, 305000241350, 30500X243874, 305000250959, 30500X254932, 305000263557 Catalog Number 110-4L: 305000221383, 30500X222566, 305000222762, 30500X223971,305000223969, 305000225078, 30500X226539, 305000226221, 30500X227058, 305000226848, 305000226749, 305000226847, 305000226871, 305000226,870, 30500X228783, 305000229450, 305000229909, 305000230871 30500X232775, 305000232199, 30500X232961, 305000233311, 305000234438, 30500X235731, 305000235181, 305000236201, 305000237368, 305000237369, 305000237799, 305000238539, 305000239102, 30500X240245, 30500X240246, 30500X240247, 305000239736, 305000239737, 30500X241365, 30500X241522, 305000241082, 30500X242170, 30500X245765, 30500X246521, 3050RX237799, 3050RX226749, 3050RX229450, 3050RX241082, 3050RX232199, 3050RY252775, 3050RX240446, 3050RY255148, 305000259553, 305000260626, 305000261835, 305000262436, 3050RX259553, 305000263146, 3050RX260626, 305000264077, 3050RX262436, 3050RX261835, 305000264175, 3050RY265351, 3050RY265766, 305000265531, 30500X266144, 3050RY266146, 30500X266522, 305000266062, 305000267297, 30500X267473, 30500X268133, 305000267784, 305000268412, 3050RX265531, 305000268916, 3050RY271046, 305000271872, 30500X272374, 305000272591, 30500X274210, 30500X274963, 305000281820 Catalog Number 110-4BC: 176044, 177018, 178013, 178898, 179757, 180139, 180955, 182279, 182908, 184079, 184687, 185284, 186277, 187293, 190452, 187780, 190451, 191891, 193414, 194074, 195161, 195608, 196987, 197492, 198189, 198512, 200069, 200490, 199609, 201039, 201595, 202423, 203565, 207077, 208532 Catalog Number 110-4HC: 179758, 180956, 183071, 184686, 186278, 187781, 190450, 192246, 194993, 195450, 198191, 199708, 201040, 201596, 202009 Catalog Number 110-4LC: 162523, 167450, 167786, 175720, 177188, 178547, 182130, 187495, 195245, 197493.
Other devices under the same FDA submission
3 devices cleared under K102875 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CAMINO® 110-4BNatus Medical Incorporated · Device, Monitoring, Intracranial Pressure8 reports
- CAMINO® 110-4BTNatus Medical Incorporated · Device, Monitoring, Intracranial Pressure1 report
- CAMINO® 110-4GNatus Medical Incorporated · Device, Monitoring, Intracranial Pressure1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗