Why it was recalled

The air pipe that delivers unfiltered ambient air from the procedure room into the water bottle on the listed FUJINON G5 & G8 type flexible GI endoscopes does not connect to the main air/water channel within the light guide (LG) connection head and, as a result, the air pipe does not receive forced flow during the Reliance EPS processing cycle.

FDA root cause category: Use error

What the company is doing

Steris sent an "URGENT PRODUCT SAFETY NOTICE" dated September 5, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.

Recall details

Recall number
Z-2263-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Aug 29, 2013
Posted by the FDA
Sep 20, 2013
Recalling firm
Steris Corporation
Quantity
266 units
Product code
NZA
FDA submissions
K040049, K102244, K110453
Distribution
Worldwide Distribution-USA (nationwide) including the states of AK, CO, FL, GA, HI, ID, IL, KS, LA, MD, MN, MO, MT, NE, NM, OH, PA, SC, UT, & WY and the countries of Australia, Canada & Spain.

Product description

Reliance Endoscope Processing System (REPS), STERIS Corporation, 5960 Heisley Road, Mentor, OH 44060-1834 USA www.steris.com. The intended use of the REPS is washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-resistant, semi-critical devices.

Code information (models, lots, serial numbers)

Model #'s: MB30001, MB30002, MB30003, MB000077, MB000078, MB000079, MB000080 and MB000081. Serial number range 3625606001 - 3608912002

Other devices under the same FDA submission

1 device cleared under K040049, K102244, K110453 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.