STERRAD 100NX Cassettes recallZ-0044-2014 · Advanced Sterilization Products · initiated Sep 18, 2013

Class II Terminated All recalls posted in October 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Advanced Sterilization Products or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Advanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.

FDA root cause category: Under Investigation by firm

What the company is doing

Advanced Sterilization Products (ASP) sent an Urgent Medical Device Recall letter on September 18, 2013, to all affected customers. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to examine their inventory and return all affected cassettes using the enclosed prepaid UPS return lave to Stericycle. Customers were asked to count their inventory and record the data on the enclosed Business Reply Card and Packing Slip. Customers were asked to provide the notice to anyone in their facility that needs to be informed. Customers were also asked to maintain a copy of the notice with the affected product. Customers with questions or suspected problems were instructed to contact Stericycle directly at 1-877-257-7119. For questions regarding this recall call 949-453-6400.

Recall details

Recall number
Z-0044-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Sep 18, 2013
Posted by the FDA
Oct 22, 2013
Recalling firm
Advanced Sterilization Products
Quantity
2,995 units
Product code
MLR
FDA submissions
K071385
Distribution
Nationwide Distribution

Product description

STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

Code information (models, lots, serial numbers)

Lot# 13F050, 13F051, 13F049

Other devices under the same FDA submission

5 devices cleared under K071385 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.