Skip to content

REF MICROMAX***microMax Handpiece***Rx Only***Manufacturer:… recallZ-0656-2014 · The Anspach Effort, Inc. · initiated Apr 25, 2013

Class II Terminated All recalls posted in January 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from The Anspach Effort, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm Anspach Effort of Palm Beach Gardens, FL initiated a voluntary removal of the Pneumatic Drill System, including the motor hand-piece, Auto Lube foot control, without a pressure relief valve (PRV) due to the hand piece's outer hose can rupture without warning if the air flow is blocked during use..

FDA root cause category: Component design/selection

What the company is doing

The firm, Anspach sent an "URGENT" Medical Devics Removal" letter dated April 24, 2013, to its customers. Customer Immediate Actions: 1. Screen your current inventory and return all Anspach Pneumatic Drill System without a pressure Relief valve (PRV) from your inventory. 2. If you have any product listed on the attached form to arrange for immediate return and replacements. 3. complete the attached Customer Reply form indicating the product being returned. 4. Upon receipt of a replacement product, equipment being returned must be received by Anspach within 10 business days of the replacement request. Should you have any queries, please contact Anspach Product Support at (800) 327-6887.

Recall details

Recall number
Z-0656-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Apr 25, 2013
Posted by the FDA
Jan 9, 2014
Recalling firm
The Anspach Effort, Inc.
Quantity
1,224
Product code
ERL
FDA submissions
K965080
Distribution
Worldwide Distribution: US (nationwide) including Puerto Rico and states of: AL, AR, AZ, CA, CO, CT, DC, DE, GA, FL, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VE, VT, WA, WI, WY, and WV. Internationally to: AT, AU, BR, BS, CA, CH, CL, CO, CZ, CR, DE, DK, DO, EG, ES, FI, FR, GB, GR, HK, HU, ID, IL, IN, IS, JP, KH, KR, LB, LT, MX, MY, NL, NO, NZ, PA, PE, PH, PL, PT, SA, SG, SY, TH, TR, TW, and ZA.

Product description

***REF MICROMAX***microMax Handpiece***Rx Only***Manufacturer: The Anspach Effort, Inc. Intended for cutting and shaping bone.

Code information (models, lots, serial numbers)

Serial Number MM900

Other devices under the same FDA submission

2 devices cleared under K965080 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 66323).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page