MAQUET Getinge Group CARDIOHELP Base Unit REF 70104.7999… recallZ-0082-2014 · Maquet Medical Systems USA · initiated Sep 12, 2013
Class II Terminated All recalls posted in October 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Maquet Medical Systems USA or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
It has come to the attention of MAQUET that in certain rare instances, the CARDIOHELP-i device may display an error message with audible alarm, indicating "battery defective." The alarm will occur-upon start up or when the main power supply is unplugged from the main power source. The human machine interface touch screen of the CARDIOHELP-i device may gradually become black due to the human machine interface (HMI) back light shutting down.
FDA root cause category: Software Manufacturing/Software Deployment
What the company is doing
MAQUET issued their customers "Urgent-Medical Device Field Correction" notices/Customer Fax Response forms dated September 16, 2013 via FedEx on September 16, 2013. The notification informs the customer of the issue with the product; how to identified affected product; and the action needed to be taken by the customer. The field correction includes service on the affected product. Authorized MAQUET Service Technicians will service the affected product to correct the issue. Customers were requested to complete and fax back (or email a scanned copy) of the enclosed response form. Customers could contact their local field representative or the MAQUET Customer Service line at 1.888.627.8383 (press option 2 followed by option 2) Monday through Friday, 8am to 6pm EST. For questions regarding this recall call 973-709-7000.
Recall details
- Recall number
- Z-0082-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 12, 2013
- Posted by the FDA
- Oct 25, 2013
- Recalling firm
- Maquet Medical Systems USA
- Quantity
- 63
- Product code
- DTQ
- FDA submissions
- K102726
- Distribution
- Nationwide Distribution including AZ, FL, CO, TX, OH, VA, MN, WI, SC, IL, ND, UT, CA, OR, NY, IA, CT, SD, MI, HI, AL, MD, KS, MO
Product description
MAQUET Getinge Group CARDIOHELP Base Unit REF 70104.7999 REF 70104.8012 Manufactured for: MAQUET Cardiopulmonary AG Made in Germany. Cardiopulmonary support system.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Part No. 70104.8012 Serial Nos. 90410440; 90410452; 90410455; 90410456; 90410457; 90410459; 90410469; 90410474; 90410480; 904104781; 90410482; 90410484; 90410498; 90410499; 90410500; 90410501; 90410502; 90410503; 90410521; 90410522; 90410524; 90410526; 90410527; 90410531; 90410533; 90410534; 90410535; 90410550; 90410551; 90410552; 90410568; 90410571; 90410572; 90410573; 90410575; 90410576; 90410577; 90410580; 90410581; 90410582; 90410584; 90410585; 90410592; 90410596; 90410599; 90410600; 90410605; 90410606; 90410607; 90410608; 90410609; 90410610; 90410611; 90410612; 90410626; 90410638; 90410640; 90410642; 90410644; 90410646; 90410647; 90410650; 90410651; 90410652; 90410653; 90410656; 90410659; 90410660; 90410661; 90410662; 90410663; 90410664; 90410665; 90410666; 90410667; 90410668; 90410670; 90410700; 90410708; 90410846; 90410750; 90410786
Other devices under the same FDA submission
1 device cleared under K102726 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA Wall holder HKH 8820Maquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page