Siemens Chemistry Calibrator (CAL) used with the ADVIA… recallZ-0365-2014 · Siemens Healthcare Diagnostics · initiated Sep 24, 2013
Class II Terminated All recalls posted in November 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Siemens Healthcare Diagnostics is conducting a field correction for Siemens Chemistry Calibrator Lots 090445B and 090445C on the ADVIA 1200, 1650, 1800, and 2400 Chemistry Systems. The System Specific Values (SSVs) for the Direct Bilirubin Assay (DBIL_2) have been reassigned for these lots. Siemens has determined that the DBIL_2 SSVs currently assigned to these lots may cause an average shift of
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens Healthcare Diagnostics sent an Urgent Medical Device Correction Letters dated September 24, 2013 and Response Forms to the US consignees via FedEx. Urgent Field Safety Notices were also distributed to customers outside the US. Customers are instructed to use the reassigned System Specific Values for the affected lots listed in the letter. Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Recall letter within thirty (30) days. The questionnaire is to be faxed to the Technical Solutions Center at 302-631-7597. On September 30, 2013, Siemens sent a revised customer notifications to correct the Part Number.
Recall details
- Recall number
- Z-0365-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 24, 2013
- Posted by the FDA
- Nov 19, 2013
- Recalling firm
- Siemens Healthcare Diagnostics
- Quantity
- Domestic: 1191 units; Foreign: 3052 units
- Product code
- CIG
- FDA submissions
- K050374
- Distribution
- Worldwide Distribution - USA Nationwide in the states of: Argentina, Austria, Belgium, Bulgaria, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, France, French Polynesia, Germany, Guadeloupe, Guatemala, Great Britain, Greece, Korea, Indonesia, Israel, Italy, Kazakhstan, Latvia, Liechstenstein, Lithuania, Luxembourg, Macedonia, Martinique, Mexico, Netherlands, Norway, Peru, Poland, Portugal, Russian Federation, Serbia, Singapore, Slovenia, Slovakia, Spain, Sweden, Switzerland, Thailand, Taiwan, Tunisia, Turkey, United Arab Emirates, Uruguay, Vatican City, and Vietnam.
Product description
Siemens Chemistry Calibrator (CAL) used with the ADVIA Chemistry Direct Bilirubin_2 Reagent, Siemens Material Number (SMN) 10312279, REF 09784096, Part Number T03-1291-62; IVD Product usage: Usage: DBIL_2: For in vitro diagnostic use in the quantitative determination of direct bilirubin in human serum and plasma on the ADVIA Chemistry systems. Such measurements are useful in the screening of liver function disorders or in the diagnosis of jaundice.
Code information (models, lots, serial numbers)
Lots 090445B (exp. 07/2014) and 090445C (exp. 10/2014)
Other devices under the same FDA submission
3 devices cleared under K050374 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Atellica CH Lithium_2 (LITH_2) 11532401Siemens Healthcare Diagnostics Inc. · Assay, porphyrin, spectrophotometry, lithium2 recalls0 reports
- ADVIA® Chemistry Calibrator 10312279Siemens Healthcare Diagnostics Inc. · CALIBRATOR, MULTI-ANALYTE MIXTURE1 recall3 reports
- Atellica CH Li 11097535Siemens Healthcare Diagnostics Inc. · Assay, porphyrin, spectrophotometry, lithium1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page